Specifications, Frequency – Atec Rohde-Schwarz-ESCI Series User Manual
Page 2
Version 03.00, June 2009
2
Rohde & Schwarz
R&S
®
ESCI/ESCI7 EMI Test Receiver
Specifications
Specifications apply under the following conditions: 15 minutes warm-up time at ambient temperature, specified environmental
conditions met, calibration cycle adhered to, and all internal automatic adjustments performed. Data without tolerances: typical values
only. Data designated 'nominal' applies to design parameters and is not assured by Rohde & Schwarz.
Frequency
R&S
®
ESCI
DC, AC coupled
9 kHz to 3 GHz
R&S
®
ESCI7
DC coupled
9 kHz to 7 GHz
Frequency range
AC coupled
1 MHz to 7 GHz
Resolution
0.01 Hz
Internal reference frequency (nominal)
standard
Aging per year
after 30 days of continuous operation
1 × 10
–6
Temperature drift
+5 °C to +45 °C
1 × 10
–6
Internal reference frequency (nominal)
R&S
®
FSP-B4 option (OCXO)
Aging per year
after 30 days of continuous operation
1 × 10
–7
Temperature drift
+5 °C to +45 °C
1 × 10
–8
External reference frequency
10
MHz
Frequency display (receiver mode)
numeric
display
Resolution
0.1
Hz
Frequency display (analyzer mode)
with marker or frequency counter
Marker resolution
span/500
Max. deviation
sweep time > 3 × auto sweep time
±(marker frequency × reference frequency
error + 0.5 % × span + 10 % × resolution
bandwidth + ½ (last digit))
Frequency counter resolution
selectable
0.1 Hz to 10 kHz
Count accuracy
S/N > 25 dB
± (marker frequency × reference
frequency error + ½ (last digit))
Display range of frequency axis
R&S
®
ESCI
0 Hz, 10 Hz to 3 GHz
R&S
®
ESCI7
0 Hz, 10 Hz to 7 GHz
Max. deviation of display range
0.1 %
f = 500 MHz, for f > 500 MHz see diagram
100 Hz
< –84 dBc (1 Hz), typ. –90 dBc (1 Hz)
1 kHz
< –100 dBc (1 Hz), typ. –108 dBc (1 Hz)
10 kHz
< –106 dBc (1 Hz), typ. –113 dBc (1 Hz)
100 kHz, span > 100 kHz
< –110 dBc (1 Hz), typ. –113 dBc (1 Hz)
1 MHz, span > 100 kHz
< –120 dBc (1 Hz), typ. –125 dBc (1 Hz)
Spectral purity, SSB phase noise
10 MHz
typ. –145 dBc (1 Hz)
Residual FM
f = 500 MHz, RBW = 1 kHz,
sweep time = 100 ms
typ. 3 Hz
-130
-120
-110
-100
-90
-80
-70
-60
100 Hz
1 kHz
10 kHz
100 kHz
1 MHz
S
S
B
p
h
ase n
o
ise (
d
B
c
(1
H
z
))
center frequency 500 MHz
center frequency 3 GHz
center frequency 7 GHz
Typical phase noise at different center frequencies