Level – Atec Rohde-Schwarz-ESCI Series User Manual
Page 4
Version 03.00, June 2009
4
Rohde & Schwarz
R&S
®
ESCI/ESCI7 EMI Test Receiver
Level
Display range
displayed average noise level (DANL)
to 30 dBm
Maximum input level
DC-coupled 0
V
DC voltage
AC-coupled 50
V
RF attenuation 0 dB
20 dBm
CW RF power
RF attenuation ≥ 10 dB
30 dBm
Pulse spectral density
RF attenuation 0 dB
97 dBμV/MHz
Max. pulse voltage
RF attenuation ≥ 10 dB, 10 μs
150 V
R&S
®
ESCI
RF attenuation ≥ 10 dB, 20 μs
10 mWs
R&S
®
ESCI7
Max. pulse energy
RF attenuation ≥ 10 dB, 10 μs
1 mWs
Intermodulation
1 dB compression of input mixer
f > 200 MHz, RF attenuation 0 dB,
preselection and preamplifier off
5 dBm, nominal
RF attenuation 0 dB, level 2 × –30 dBm, ∆f > 5 × RBW or 10 kHz, whichever is larger
without preselection, without preamplifier
R&S
®
ESCI, R&S
®
ESCI7
20 MHz to 200 MHz
> 5 dBm
200 MHz to 3 GHz
> 7 dBm, typ. 10 dBm
R&S
®
ESCI7
3 GHz to 7 GHz
> 10 dBm, typ. 15 dBm
with preselection, without preamplifier
R&S
®
ESCI, R&S
®
ESCI7
20 MHz to 200 MHz
> 0 dBm
200 MHz to 3 GHz
> 2 dBm, typ. 5 dBm
R&S
®
ESCI7
3 GHz to 7 GHz
> 10 dBm, typ. 15 dBm
with preselection, with preamplifier
R&S
®
ESCI, R&S
®
ESCI7
20 MHz to 200 MHz
> –20 dBm
200 MHz to 3 GHz
> –18 dBm, typ. –15 dBm
R&S
®
ESCI7
Third-order intercept (TOI)
3 GHz to 7 GHz
> –10 dBm, typ. –5 dBm
RF attenuation 0 dB, level –10 dBm, without preselection, without preamplifier
R&S
®
ESCI, R&S
®
ESCI7
< 100 MHz
typ. 25 dBm
100 MHz to 1.5 GHz
typ. 35 dBm
R&S
®
ESCI7
1.5 GHz to 3.5 GHz
typ. 70 dBm
RF attenuation 0 dB, level –15 dBm, with preselection, without preamplifier
R&S
®
ESCI, R&S
®
ESCI7
4 MHz to 100 MHz
> 40 dBm
100 MHz to 1.5 GHz
> 50 dBm
R&S
®
ESCI7
1.5 GHz to 3.5 GHz
typ. 70 dBm
RF attenuation 0 dB, level –35 dBm, with preselection, with preamplifier
R&S
®
ESCI, R&S
®
ESCI7
4 MHz to 100 MHz
> 25 dBm
100 MHz to 1.5 GHz
> 35 dBm
R&S
®
ESCI7
Second harmonic intercept (SHI)
1.5 GHz to 3.5 GHz
typ. 10 dBm