Tos9200 series, Hipot tester with insulation resistance test, General specifi cations – Atec Kikusui-TOS9201 User Manual

Page 9

background image

12 Kikusui Electronics Corporation

TOS9200 SERIES

Hipot Tester with Insulation Resistance Test

Item

TOS9200

TOS9201

Judgement function

Judgement method/action

• The PASS signal is output at the timing preset on PASS HOLD. If HOLD is set, the PASS signal is output continuously until
the STOP signal is input.
• The UPPER FAIL signal and the LOWER FAIL signal are output continuously until the STOP signal is input.

• The FAIL and PASS buzzer volumes are adjustable. However, they cannot be adjusted individually, as they are set in common.

Setting range for the upper resistance (UPPER)

0.01 M

Ω to 9.99 GΩ [Below the maximum rated current]

Setting range for the lower resistance (LOWER)

0.01 M

Ω to 9.99 GΩ [Below the maximum rated current]

Judgement accuracy
For both UPPER and LOWER

Time

Setting range for the voltage rise time (RISE TIME)

0.1 s to 200 s

Setting range for the test time (TEST TIME)

0.5 s to 999 s With the TIMER OFF function

Setting range for the judgement wait time (WAIT TIME)

0.3 s to 10 s [RISE TIME + TEST TIME > WAIT TIME]

Accuracy

± (100 ppm + 20 ms)

*

5 When the GND is set to GUARD, current measurement is disabled if the part of the DUT connected to the LOW terminal is grounded, which poses extreme danger. Never ground the DUT.

In ordinary operation, set the GND to LOW.

Judgement

Judgement method

Display

Buzzer

SIGNAL I/O

UPPER FAIL

When the tester detects a resistance exceeding the upper cutoff resistance,

it cuts off the output and makes an UPPER FAIL judgement. However,

no judgement is made during a voltage rise time (RISE TIME).

The FAIL

LED lights up.

Displayed

on the LCD

ON

Outputs the

U FAIL signal

LOWER FAIL

When the tester detects a resistance below the lower cutoff resistance,

it cuts off the output and makes a LOWER FAIL judgement. However,

no judgement is made until the judgement wait time (WAIT TIME)

has elapsed.

The FAIL

LED lights up.

Displayed

on the LCD

ON

Outputs the

L FAIL signal

PASS

When the preset time has elapsed without any abnormalities,

the tester cuts off the output and makes a PASS judgement.

The PASS

LED lights up.

Displayed

on the LCD

ON

Outputs the

PASS signal

Judgement current

50 nA

≤ i ≤ 100 nA

100 nA < i

≤ 200 nA 200nA < i ≤ 1 μA

1

μA < i ≤ 1 mA

UPPER, LOWER

0.01

≤ R < 10.0 MΩ

––

––

––

± (2 % of setting + 3digit)

10.0

≤ R < 50.0 MΩ

––

––

± (5 % of setting + 5digit)

± (2 % of setting + 3digit)

50.0

≤ R < 100 MΩ

––

––

± (5 % of setting + 5digit)

± (2 % of setting + 3digit)

100 M

Ω ≤ R < 200 MΩ

––

± (10 % of setting + 5digit)

± (5 % of setting + 5digit)

± (2 % of setting + 3digit)

200 M

Ω ≤ R < 500 MΩ

± (20 % of setting + 5digit)

± (10 % of setting + 5digit)

± (5 % of setting + 5digit)

± (2 % of setting + 3digit)

500 M

Ω ≤ R < 1.00 GΩ

± (20 % of setting + 5digit)

± (10 % of setting + 5digit)

± (5 % of setting + 5digit)

± (2 % of setting + 3digit)

1.00 G

Ω ≤ R < 2.00 GΩ

± (20 % of setting + 10digit)

± (10 % of setting + 5digit)

± (5 % of setting + 5digit)

––

2.00 G

Ω ≤ R < 5.00 GΩ

± (20 % of setting + 20digit) ± (10 % of setting + 10digit)

± (5 % of setting + 5digit)

––

5.00 G

Ω ≤ R < 10.0 GΩ

± (20 % of setting + 20digit) ± (10 % of setting + 10digit)

––

––

Judgement current = test voltage/(UPPER,LOWER)
[In the humidity range of 20 %rh to 70 %rh (no codensation), with no disturbance such as swinging of the test leadwire]
[In LOWER judgement, at least 0.5 s is necessary for testing after the WAIT TIME has elapsed. In LOWER judgement
for 200 nA or lower, a wait time of at least 1.0 s is necessary.]

General Specifi cations

Item

TOS9200

TOS9201

Environment

Installation location

Indoors at an altitude of up to 2000 m

Warranty range

Temperature

5

°C to 35 °C

Humidity

20 %rh to 80 %rh (No condensation)

Operating range

Temperature

0

°C to 40 °C

Humidity

20 %rh to 80 %rh (No condensation)

Storage range

Temperature

-20

°C to 70 °C

Humidity

90 %rh or less (No condensation)

Power requirements

Nominal voltage range (Allowable voltage range)

100 V to 120 V AC / 200 V to 240 V AC (85 V to 130 V AC / 170 V to 250 V AC) Selectable

Power consumption

Using no load (READY)

100 VA or less

Using the rated load

Maximum of 800 VA

Allowable frequency range

47 Hz to 63 Hz

Insulation resistance

30 M

Ω or more (500 V DC) [between the AC LINE and chassis]

Hipot

1390 V AC, 2 seconds, 20 mA or less [between the AC LINE and chassis]

Ground bond

25 A AC/0.1

Ω or less

Electromagnetic compatibility (EMC) (*6)

Conforms to the requirements of the following directive and standard.

EMC Directive 89/336/EEC, EN61326, EN61000-3-2, EN61000-3-3

Under following conditions

1. Used test leadwire TL01-TOS which is supplied. 2. No discharge occurs at outside of the tester.

3. Used the shielded cable which length is less than three meters when the SIGNAL I/O is used.

Safety (*6,7)

Conforms to the requirements of the following directive and standard.
Low Voltage Directive 73/23/EEC , EN61010-1, Class I, Pollution degree 2

Dimensions (maximum)

430 (455) W x 132 (150) H x 370 (440) D mm

Weight

Approx. 19 kg