beautypg.com

Tos9200 series, Hipot tester with insulation resistance test, External dimensional diagrams – Atec Kikusui-TOS9201 User Manual

Page 10: Electrical performance

background image

13

Kikusui Electronics Corporation

TOS9200 SERIES

Hipot Tester with Insulation Resistance Test

Item

TOS9200

TOS9201

Accessory

AC Power cable

1 pc.

High-voltage test lead wire TL01-TOS (1.5 m)

1 set

Interlock jumper

1 pc.

High-Voltage Danger seal

1 sheet

Fuse

1 pc.

Operation Manual

Operation Manual for Tester: 1 copy, Operation for GPIB/RS-232C Interface: 1 copy

*6

Only on models that have CE marking on the panel. Not applicable to custom order models.

*7 This instrument is a Class I equipment. Be sure to ground the protective conductor terminal of the instrument. The safety of the instrument is not guaranteed unless the instrument is grounded properly.

Electrical performance

Item

TOS9200

TOS9201

Maximum rating
voltage

AC

5.0 kV

DC

6.0 kV

Number of channels

4 (Each channel is settable to HIGH, LOW, or OPEN.)

Maximum number of scanners connected

4 scanners,Channel numbers are determined in order of connection to the TOS9200/9201 tester.

1 st scanner CH1 to CH4 2 nd scanner CH5 to CH8 3 rd scanner CH9 to CH12 4 th scanner CH13 to CH16

Contact check function

None (*1)

Provided

Lamps and LEDs

POWER

Lights as it is interlocked with the POWER switch of the TOS9200/9201 tester

DANGER

Lights as it is interlocked with the DANGER lamp of the TOS9200/9201 tester

CHANNEL

Lights during a test at each channel HIGH: red; LOW: green; Under contact check: orange

Power requirements

Nominal voltage range (allowable voltage range)

100 V to 120 V AC/200 V to 240 V AC (85 V to 132 V AC/170 V to 250 V AC) Automatic switching

Power consumption

In READY state

Approx. 12 VA

During test

40 VA maximum

Allowable frequency range

47 Hz to 63 Hz

Insulation resistance

30 M

Ω or more (500 V DC) [between the AC LINE and chassis]

Hipot

1390 V AC, 2 seconds, 10 mA or less [between the AC LINE and chassis]

Ground bond

25 A AC/0.1

Ω or less

Electromagnetic compatibility (EMC) (*2)

Conforms to the requirements of the following directive and standard.

EMC Directive 89/336/EEC, EN61326, EN61000-3-2, EN61000-3-3

Under following conditions

1. Used test leadwire TL07-TOS which is supplied. 2. No discharge occurs at outside of the tester.

3. Used the shielded cable which length is less than three meters when the SIGNAL I/O is used.

Safety (*2,3)

Conforms to the requirements of the following directive and standard.

Low Voltage Directive 73/23/EEC, EN61010-1, Class I, Pollution degree 2

Environment

Installation location

Indoors and at altitudes up to 2000 m

Warranty range

Temperature

5 °C to 35 °C

Humidity

20 %rh to 80 %rh (no condensation)

Operating range

Temperature

0 °C to 40 °C

Humidity

20 %rh to 80 %rh (no condensation)

Storage range

Temperature

-20 °C to 70 °C

Humidity

90 %rh or less (no condensation)

Dimensions

430(435)W

✕ 88(105)H ✕ 370(415) Dmm

Weight

Approx. 6.5 kg

Accessories

AC power cable

1 pc.

High-voltage test leadwires, red

4 pc. (1.5 m each)

8 pc. (1.5 m each)

High-voltage leads for parallel connection

1 set (0.5 m each)

Interface cable

1 pc.(0.5 m)

Channel-indication stickers

For the panel face: 1 sheet; for the test leadwires: 1

“HIGH VOLTAGE, DANGER” stickers

2 sheets

Fuses

2 pc. (including a spare contained in the fuse holder)

Operation Manual

1 copy

*2 Only on models that have CE marking on the panel. Not applicable to custom order models.
*3 This instrument is a Class I equipment. Be sure to ground the protective conductor terminal of the instrument. The safety of the instrument is not guaranteed unless the instrument is grounded properly.
[Measurement accuracy achieved when the scanner and the TOS9220/9201 tester are connected]
In an AC hipot test, a current of approx. 22 µA/kV fl ows per scanner due to stray capacitance in the scanner in comparison with use of the TOS9220/9201 tester alone. Note that
this current may contribute to errors in current measurements conducted by the TOS9220/9201 tester.

External dimensional diagrams

MAX455

430

MAX10

MAX150

132

MAX440

370

MAX

30

430

MAX435

MAX105

88

370

MAX5

MAX415

Unit: mm