Internal atomic clock (option 1), High accuracy power meter (option 19) – Atec Anritsu-MT2720T User Manual
Page 4
PN: 11410-00646 Rev. H
MS2720T TDS
Spectrum Analyzer Specifications
Spectrum Master™ MS2720T
Spectrum Analyzer
(continued)
Third-Order Intercept (TOI)
(–20 dBm tones 100 kHz apart, 0 dB Attenuation Preamp OFF)
2.4 GHz
+14 dBm minimum
50 MHz to 20 GHz
+20 dBm typical
> 20 GHz to 32 GHz
+15 dBm typical
> 32 GHz to 43 GHz
+20 dBm typical
P1dB
< 4 GHz
5 dBm typical
4 GHz to 20 GHz
12 dBm typical
> 20 GHz to 32 GHz
7 dBm typical
> 32 GHz to 43 GHz
12 dBm typical
Second Harmonic Distortion
(0 dB input attenuation, –30 dBm input)
50 MHz
–54 dBc maximum
VSWR
(> 10 dB input attenuation)
9 GHz Instruments
< 4 GHz
1.5:1 typical
4 GHz to 8 GHz
1.8:1 typical
13 GHz to 43 GHz Instruments
< 20 GHz
1.5:1 typical
20 GHz to 43 GHz
2.0:1 typical
Internal Atomic Clock (Option 1)
Frequency Accuracy
1 x 10
-9
+ aging
Aging Rate
1 x 10
-9
per year
Initial Warmup
10 minutes to meet the accuracy specification
High Accuracy Power Meter (Option 19)
(Requires external USB Power Sensor)
Amplitude
Maximum, Minimum, Offset, Relative On/Off, Units, Auto Scale
Average
# of Running Averages, Max Hold
Zero/Cal
Zero On/Off, Cal Factor (Center Frequency, Signal Standard)
Limits
Limit On/Off, Limit Upper/Lower
Power Sensor Model
PSN50
MA24105A
MA24106A
MA24108A/18A/26A
Description
High Accuracy
RF Power Sensor
Inline High
Power Sensor
High Accuracy
RF Power Sensor
Microwave USB
Power Sensor
Frequency Range
50 MHz to 6 GHz
350 MHz to 4 GHz
50 MHz to 6 GHz
10 MHz to 8/18/26 GHz
Connector
Type N(m), 50 Ω
Type N(f), 50 Ω
Type N(m), 50 Ω
Type N(m), 50 Ω
(8/18 GHz)
Type K(m), 50 Ω
(26 GHz)
Dynamic Range
–30 dBm to +20 dBm
(0.001 mW to 100 mW)
+3 dBm to +51.76 dBm
(2 mW to 150 W)
–40 dBm to +23 dBm
(0.1 µW to 200 mW)
–40 dBm to +20 dBm
(0.1 µW to 100 mW)
VBW
100 Hz
100 Hz
100 Hz
50 kHz
Measurand
True-RMS
True-RMS
True-RMS
True-RMS, Slot Power,
Burst Average Power
Measurement Uncertainty
± 0.16 dB
1
± 0.17 dB
2
± 0.16 dB
1
± 0.18 dB
3
Data sheet
(for complete specifications)
11410-00414
11410-00621
11410-00424
11410-00504
Notes:
1. Total RSS measurement uncertainty (0 ºC to 50 ºC) for power measurements of a CW signal greater than –20 dBm with
zero mismatch errors.
2. Expanded uncertainty with K=2 for power measurements of a CW signal greater than +20 dBm with a matched load.
Measurement results referenced to the input side of the sensor.
3. Expanded uncertainty with K=2 for power measurements of a CW signal greater than –20 dBm with zero mismatch
errors.