2 of 24

PN: 11410-00646 Rev. H

MS2720T TDS

Spectrum Analyzer Specifications

Spectrum Master™ MS2720T

Spectrum Analyzer

All specifications and characteristics apply to Revision 2 instruments under the following conditions, unless otherwise stated: 1) After 5 minutes of warm-up time, where the

instrument is left in the ON state; 2) Apply when using internal reference and performance sweep mode; 3) Subject to change without notice; 4) Typical performance is the

measured performance of an average unit; 5) Recommended calibration cycle is 12 months.

Measurements

Smart Measurements

Field Strength (dBm/m

2

, dBW/m

2

, V/m, A/m, Watt/m

2

, Watt/cm

2

, or dBmV/m)

Occupied Bandwidth (measures 99 % to 1 % power channel of a signal)
Channel Power (measures the total power in a specified bandwidth)
ACPR (adjacent channel power ratio)
AM/FM/SSB Demodulation (AM, wide/narrow FM, upper/lower SSB), (audio out only)
C/I (carrier-to-interference ratio)
Emission Mask (recall limit lines as emission mask)

Setup Parameters

Frequency

Center/Start/Stop, Span, Frequency Step, Frequency Offset, Signal Standard, Channel #

Amplitude

Reference Level (RL), Scale, Attenuation Auto/Level, RL Offset, Pre-Amp On/Off, Detection

Span

Span, Span Up/Down (1-2-5), Full Span, Zero Span, Last Span

Bandwidth

RBW, Auto RBW, VBW, Auto VBW, RBW/VBW Ratio, Span/RBW Ratio

Sweep Functions

Sweep

Single/Continuous, Manual Trigger, Reset, Detection, Minimum Sweep Time, Trigger Type

Sweep Mode

Fast (100x Performance), Performance, No FFT, Burst Detect (1000x Fast in 15 MHz span)

Detection

Peak, RMS/Avg, Negative, Sample, Quasi-peak

Triggers

Free Run, External, Video, Manual, IF Power

Trigger Position

Delay, Level, Slope, Hysteresis, Holdoff, Force Trigger Once

Trace Functions

Traces

Up to three Traces (A, B, C), View/Blank, Write/Hold, Trace A/B/C Operations

Trace A Operations

Normal, Max Hold, Min Hold, Average, # of Averages, (always the live trace)

Trace B Operations

A → B, B ← → C, Max Hold, Min Hold

Trace C Operations

A → C, B ← → C, Max Hold, Min Hold, A – B → C, B – A → C, Relative Reference (dB), Scale

Marker Functions

Markers

Markers 1-6 each with a Delta Marker, or Marker 1 Reference with Six Delta Markers,

Marker Table (On/Off/Large), All Markers Off

Marker Types

Style (Fixed/Tracking), Noise Marker, Frequency Counter Marker

Marker Auto-Position

Peak Search, Next Peak (Right/Left), Peak Threshold %, Set Marker to Channel,

Marker Frequency to Center, Delta Marker to Span, Marker to Reference Level

Marker Table

1-6 markers frequency and amplitude, plus delta markers frequency offset and amplitude

Limit Line Functions

Limit Lines

Upper/Lower, On/Off, Edit, Move, Envelope, Advanced, Limit Alarm, Default Limit

Limit Line Edit

Frequency, Amplitude, Add Point, Add Vertical, Delete Point, Next Point Left/Right

Limit Line Move

To Current Center Frequency, By dB or Hz, To Marker 1, Offset from Marker 1

Limit Line Envelope

Create Envelope, Update Amplitude, Number of Points (41), Offset, Shape Square/Slope

Limit Line Advanced

Type (Absolute/Relative), Mirror, Save/Recall

Frequency

Frequency Range

(usable to 0 Hz)

MS2720T-0709

9 kHz to 9 GHz

MS2720T-0713

9 kHz to 13 GHz

MS2720T-0720

9 kHz to 20 GHz

MS2720T-0732

9 kHz to 32 GHz

MS2720T-0743

9 kHz to 43 GHz

Tuning Resolution

1 Hz

Frequency Reference

Aging: ± 1.0 x 10

–6

per 10 years

Accuracy: ± 0.3 x 10

–6

(25 °C ± 25 °C) plus aging

(see Options 1 and 31 for improved frequency reference aging and accuracy)

Auto-sensing External Frequency

Reference (MHz)

1, 1.2288, 1.544, 2.048, 2.4576, 4.8, 4.9152, 5, 9.8304, 10, 13, 19.6608

Sweep Time

10 µs to 600 s in zero span

Sweep Time Accuracy

± 2 % in zero span

Bandwidth

Resolution Bandwidth (RBW)

1 Hz to 10 MHz in 1–3 sequence ± 10 % (–3 dB bandwidth)

Video Bandwidth (VBW)

1 Hz to 10 MHz in 1–3 sequence (–3 dB bandwidth)

RBW with Quasi-Peak Detection

200 Hz, 9 kHz, 120 kHz (–6 dB bandwidth)

VBW with Quasi-Peak Detection

Auto VBW is On, RBW/VBW = 1

VBW/Average Type

Linear/Log