Spectrum analyzer – Atec Anritsu-MT2720T User Manual
Page 3
Spectrum Master™ MS2720T
Spectrum Analyzer Specifications
MS2720T TDS
PN: 11410-00646 Rev. H
3 of 24
Spectrum Analyzer
(continued)
Spectral Purity – SSB Phase Noise at 1 GHz
Amplitude Ranges
Dynamic Range
>106 dB minimum at 2.4 GHz, 2/3 (TOI-DANL) in 1 Hz RBW
Measurement Range
DANL to +30 dBm
Display Range
1 to 15 dB/div in 1 dB steps, ten divisions displayed
Reference Level Range
–120 dBm to +30 dBm
Attenuator Resolution
0 to 65 dB, 5.0 dB steps
Amplitude Units
Log Scale Modes: dBm, dBV, dBmV, dBμV
Linear Scale Modes: nV, μV, mV, V, kV, nW, μW, mW, W, kW
Maximum Continuous Input
+30 dBm Peak typical, ± 50 VDC (≥ 10 dB Attenuation)
+23 dBm Peak typical, ± 50 VDC (< 10 dB Attenuation)
+13 dBm Peak typical, ± 50 VDC (Preamp = ON)
Amplitude Accuracy
Displayed Average Noise Level (DANL)
(RMS detection, VBW/Avg type = Log, Ref Level = –20 dBm for Preamp Off and –50 dBm for
Preamp On, Performance Sweep Mode)
Spurs
Offset
9 GHz Instrument
13 GHz to 43 GHz Instruments
10 kHz
100 kHz
1 MHz
10 MHz
Maximum
–108 dBc/Hz
–110 dBc/Hz
–118 dBc/Hz
–129 dBc/Hz
Typical
–112 dBc/Hz
–115 dBc/Hz
–123 dBc/Hz
–133 dBc/Hz
Maximum
–102 dBc/Hz
–106 dBc/Hz
–111 dBc/Hz
–123 dBc/Hz
Typical
–106 dBc/Hz
–110 dBc/Hz
–116 dBc/Hz
–129 dBc/Hz
20 °C to 30 °C
(after 30 minute warm-up)
–10 °C to 55 °C
(after 60 minute warm-up)
9 GHz Instrument
9 kHz to 100 kHz
a
100 kHz to 7 GHz
> 7 GHz to 9 GHz
a. Values below 100 kHz are with the preamplifier turned off.
Maximum
–
± 1.3 dB
± 1.8 dB
Typical
± 1.0 dB
± 0.5 dB
± 0.5 dB
Maximum
–
± 2.3 dB
± 2.8 dB
Typical
± 1.0 dB
± 0.5 dB
± 0.5 dB
20 GHz Instruments
100 kHz to 13 GHz
> 13 GHz to 18 GHz
> 18 GHz to 20 GHz
± 1.3 dB
± 2.3 dB
–
± 0.5 dB
± 0.5 dB
± 1.0 dB
± 2.3 dB
± 3.3 dB
–
± 05 dB
± 0.5 dB
± 1.0 dB
43 GHz Instruments
> 100 kHz to 13 GHz
> 13 GHz to 40 GHz
> 40 GHz to 43 GHz
± 1.3 dB
± 2.3 dB
–
± 0.5 dB
± 0.5 dB
± 1.0 dB
± 2.3 dB
± 3.3 dB
–
± 0.5 dB
± 0.5 dB
± 1.0 dB
Preamp = Off
Preamp = On
9 GHz Instrument
10 MHz to 3 GHz
> 3 GHz to 8 GHz
> 8 GHz to 9 GHz
Maximum
–146 dBm
–140 dBm
–
Typical
–149 dBm
–143 dBm
–142 dBm
Maximum
–160 dBm
–152 dBm
–
Typical
–163 dBm
–155 dBm
–155 dBm
13 to 43 GHz Instruments
10 MHz to 4 GHz
> 4 GHz to 9 GHz
> 9 GHz to 13 GHz
–145 dBm
–142 dBm
–136 dBm
–148 dBm
–145 dBm
–139 dBm
–161 dBm
–159 dBm
–156 dBm
–164 dBm
–162 dBm
–159 dBm
20 GHz Instrument
> 13 GHz to 20 GHz
–138 dBm
–141 dBm
–157 dBm
–160 dBm
32 to 43 GHz Instruments
> 13 GHz to 32 GHz
> 32 GHz to 40 GHz
> 40 GHz to 43 GHz
–135 dBm
–127 dBm
–
–138 dBm
–130 dBm
–130 dBm
–154 dBm
–148 dBm
–
–157 dBm
–151 dBm
–151 dBm
(RF input terminated, 0 dB input attenuation)
Residual Spurs
< 13 GHz
13 to 20 GHz
> 20 to 32 GHz
> 32 to 43 GHz
Preamp = Off
–90 dBm, maximum
–85 dBm, maximum
–80 dBm, maximum
–80 dBm, maximum
Preamp = On
–100 dBm, maximum
–100 dBm, maximum
–100 dBm, maximum
–95 dBm, maximum
Input-Related Spurious
–60 dBc, –70 dBc typical (0 dB attenuation, –30 dBm input, span < 1.7 GHz)