9

Table 1-26. EMC and safety
Description

Supplemental information

EMC

• European council directive 89 / 336 /

EEC, 92 / 31 / EEC, 93 / 68 / EEC
IEC 61326 - 1: 1997 +A1: 1998 +A2:
2000/EN 61326 - 1: 1997 +A1: 1998
+A2: 2001

• CISPR 11: 1997 +A1: 1999 / EN 55011:

1998 +A1: 1999

• Group 1, Class A
• IEC 61000 - 4-2:1995 +A1: 1998 / EN

61000 - 4-2: 1995 +A1: 1998

• 4 kV CD / 8 kV AD
• IEC 61000 - 4-3: 1995 +A1: 1998 / EN

61000 - 4-3: 1996 +A1: 1998

• 3 V / m, 80 - 1000 MHz, 80% AM
• IEC 61000 - 4-4: 1995 / EN 61000 - 4-4:

1995

• 1 kV power / 0.5 kV signal
• IEC 61000 - 4-5: 1995 / EN 61000 - 4-5:

1995

• 0.5 kV normal / 1 kV common
• IEC 61000 - 4-6: 1996 / EN 61000 - 4-6:

1996

• 3 V, 0.15-80 MHz, 80% AM
• IEC 61000 - 4-11: 1994 / EN 61000 - 4-11:

1994

• 100% 1 cycle European council directive

ICES/NMB-001

This ISM device complies with Canadian
ICES-001:1998
AS/NZS 2064.1/2 Group 1, Class A

Safety

European council directive 73/23/EEC,
93/68/EEC
IEC 61010-1:2001/EN 61010-1:2001
Measurement category I, pollution degree 2,
indoor use
IEC60825-1:1994 Class 1 LED

CAN/CSA C22.2 No. 1010.1-92

Table 1-27. Analyzer environment and dimensions
Description Supplemental

information

Operating environment
Temperature

+10 °C to +40 °C

Humidity

20% to 80% at wet bulb temperature
< +29 °C (non-condensing)

Altitude

0 to 2,000 m (0 to 6,561 feet)

Vibration

0.5 G maximum, 5 Hz to 500 Hz

Non-operating storage environment
Temperature

-10 °C to +60 °C

Humidity

20% to 90% at wet bulb temperature
< +40 °C (non-condensing)

Altitude

0 to 4,572 m (0 to 15,000 feet)

Vibration

0.5 G maximum, 5 Hz to 500 Hz

Dimensions

See Figures 1-5 through 1-7

Weight (net)

21 kg