Table 1-15. Narrow band transient (frequency bandwidth = 3.125 kHz)
Measurement time (sec)
100 m
200 m
500 m
1
2
5
10
Frequency resolution (Hz rms)
0.0095
0.0095
0.0034
0.0012
0.0004
0.0004
0.0004
Time resolution (µsec)
81.92
163.84
409.6
819.2
1638.4
4096
10240
Number of point
1222
1222
1222
1222
1222
1222
978
Table 1-16. Narrow band transient (frequency bandwidth = 25 kHz)
Measurement time (sec)
10 m
20 m
50 m
100 m
200 m
500 m
1
2
5
10
Frequency resolution (Hz rms)
0.22
0.22
0.076
0.027
0.0095
0.0095
0.0095
0.0095
0.0095
0.0095
Time resolution (µsec)
10.24
20.48
51.2
102.4
204.8
512
1280
2560
6400
12800
Number of point
978
978
978
978
978
978
783
783
783
783
Table1-17. Narrow band transient (frequency bandwidth = 200 kHz)
Measurement time (sec)
1 m
2 m
5 m
10 m
20 m
50 m
100 m
200 m
500 m
1
2
5
10
Frequency resolution (Hz rms)
4.9
4.9
1.7
0.6
0.2
0.2
0.2
0.2
0.2
0.2
0.2
0.2
0.2
Time resolution (µsec)
1.28
2.56
6.4
12.8
25.6
64
160
320
800
1600
3200
8000
16000
Number of point
783
783
783
783
783
783
626
626
626
626
626
626
626
Table1-18. Narrow band transient (frequency bandwidth = 1.6 MHz)
Measurement time (sec)
100 µ 200 µ 500 µ
1 m
2 m
5 m
10 m
20 m 50 m
100 m 200 m 500 m
1
2
5
10
Frequency resolution (Hz rms)
110
110
110
39
21
14
5
5
5
5
5
5
5
5
5
5
Time resolution (µsec)
0.16
0.32
0.8
0.8
1.6
4
8
20
80
160
320
800
1600 3200 8000 1600
Number of point
626
626
626
1251 1251 1251 1251
1001
626
626
626
626
626
626
626
626
Table1-19. Narrow band transient (frequency bandwidth = 25.6 MHz)
Measurement time (sec)
10 µ 20 µ 50 µ 100 µ 200 µ 500 µ 1 m 2 m
5 m 10 m 20 m 50 m 100 m 200 m 500 m
1
2
5
Frequency resolution (Hz rms) 7 k
7 k
7 k
7 k
3 k
884
313
313
313
313
313
313
313
313
313
313 312.5 312.5
Time resolution (µsec)
0.01 0.02 0.05
0.1
0.2
0.5
1
2
6.25 12.5
25
62.5
125
250
625 1250 2500 6250
Number of point
1001 1001 1001 1001 1001 1001 1001 1001 801
801
801
801
801
801
801
801
801
801
Table 20. Spectrum monitor measurement capabilities
Spectrum monitor
Number of trace
1 data trace and 1 memory trace per measurement trace
Data formats
dBm, dBV, watt, volt, dBm / Hz, dBV / Hz, watt / Hz, volt /
√
Hz
Measurement trigger
Set to continuous, hold, or single, sweep with internal, external, manual, or bus trigger