Frequency, rf power, dc current measurements – Atec Agilent-E5052A User Manual

Page 5

background image

5

Table 1-5. Typical measurement time (sec) for phase noise
measurement

1

Table 1-6. Measurement capabilities
Phase noise
Measurement parameters

SSB phase noise, spurious, integrated
phase noise, rms noise, rms jitter,
residual FM

Number of trace

1 data trace and 1 memory trace

Data formats

dBc/Hz (SSB phase noise), dBc
(spurious), rms degree, rms radian
(rms noise), sec (rms jitter), Hz rms
(residual FM)

Measurement trigger

Set to continuous, hold, or single,
sweep with internal, external,
manual, or bus trigger

Frequency, RF Power, DC Current Measurements

Table 1-7. Frequency measurement
Description Specifications

Frequency range

10 M to 1.5 GHz
300 M to 7 GHz

Resolution

10 Hz, 1 kHz, 64 kHz

Internal time base stability

± 5 ppm

Accuracy

± (resolution + time base accuracy)

Table 1-8. RF power measurement
Description Specifications

Frequency range band

10 M to 1.5 GHz
300 M to 7 GHz

Input level
10 MHz to 30 MHz

-15 dBm to +20 dBm

30 MHz to 7 GHz

-20 dBm to +20 dBm

Resolution

0.01 dB

Accuracy (peak voltage response)
30 MHz to 3 GHz, > -10 dBm

± 0.5 dB

Other than the above

± 1 dB

Table 1-9. DC current measurement
Description Specifications

Current range

0 to 80 mA

Resolution

10 µA

Accuracy

± (0.2% of reading + 160 µA)

Table 1-10. Frequency, RF power, DC current measurements (standard)

Frequency, power, and DC current (for standard)

Measurement

Analyzer mode:

parameters

Frequency versus DC control voltage, dF / dV
control (tuning sensitivity)
Frequency versus DC power voltage
(frequency pushing), dF / dV power
RF power versus DC control voltage
RF power versus DC power voltage
DC current (at DC power port only)
versus DC control voltage
DC current (at DC power port only)
versus DC power voltage
Tester mode:
Frequency, power, and DC current
(at DC power port)

Number of points 2 to 1001
Data formats

Frequency: Hz, dHz, %, ppm, Hz/V
RF Power: dBm
DC Current: A

Measurement
trigger

Set to continuous, hold, or single, sweep with
internal, external, manual, or bus trigger

Table 1-11. Frequency, power, and DC current (Option E5052A-011)
Frequency, power, and DC current (Option E5052A-011)
Measurement

Frequency, power and DC current (at DC power

parameters

port) (numerical display only)

Data format

Frequency: Hz, dHz, %, ppm, Hz/V
RF Power: dBm
DC Current: A

Measurement

Set to continuous, hold, or single with internal,

trigger

external, manual, or bus trigger

1

10

100

1k

100 k

10.9

2.7

0.34

0.04

1 M

10.7

2.7

0.34

0.04

10 M

11.1

2.8

0.35

0.04

40 M

17.7

4.4

0.56

0.07

Stop frequency (Hz)

Start frequency (Hz)

1. Measurement time (sec) = 0.2 + the above value x number of correlation when applying
cross-correlation function (standard ONLY). For E5052-011, number of correlation = 1.