Frequency, rf power, dc current measurements – Atec Agilent-E5052A User Manual
Page 5
5
Table 1-5. Typical measurement time (sec) for phase noise
measurement
1
Table 1-6. Measurement capabilities
Phase noise
Measurement parameters
SSB phase noise, spurious, integrated
phase noise, rms noise, rms jitter,
residual FM
Number of trace
1 data trace and 1 memory trace
Data formats
dBc/Hz (SSB phase noise), dBc
(spurious), rms degree, rms radian
(rms noise), sec (rms jitter), Hz rms
(residual FM)
Measurement trigger
Set to continuous, hold, or single,
sweep with internal, external,
manual, or bus trigger
Frequency, RF Power, DC Current Measurements
Table 1-7. Frequency measurement
Description Specifications
Frequency range
10 M to 1.5 GHz
300 M to 7 GHz
Resolution
10 Hz, 1 kHz, 64 kHz
Internal time base stability
± 5 ppm
Accuracy
± (resolution + time base accuracy)
Table 1-8. RF power measurement
Description Specifications
Frequency range band
10 M to 1.5 GHz
300 M to 7 GHz
Input level
10 MHz to 30 MHz
-15 dBm to +20 dBm
30 MHz to 7 GHz
-20 dBm to +20 dBm
Resolution
0.01 dB
Accuracy (peak voltage response)
30 MHz to 3 GHz, > -10 dBm
± 0.5 dB
Other than the above
± 1 dB
Table 1-9. DC current measurement
Description Specifications
Current range
0 to 80 mA
Resolution
10 µA
Accuracy
± (0.2% of reading + 160 µA)
Table 1-10. Frequency, RF power, DC current measurements (standard)
Frequency, power, and DC current (for standard)
Measurement
Analyzer mode:
parameters
Frequency versus DC control voltage, dF / dV
control (tuning sensitivity)
Frequency versus DC power voltage
(frequency pushing), dF / dV power
RF power versus DC control voltage
RF power versus DC power voltage
DC current (at DC power port only)
versus DC control voltage
DC current (at DC power port only)
versus DC power voltage
Tester mode:
Frequency, power, and DC current
(at DC power port)
Number of points 2 to 1001
Data formats
Frequency: Hz, dHz, %, ppm, Hz/V
RF Power: dBm
DC Current: A
Measurement
trigger
Set to continuous, hold, or single, sweep with
internal, external, manual, or bus trigger
Table 1-11. Frequency, power, and DC current (Option E5052A-011)
Frequency, power, and DC current (Option E5052A-011)
Measurement
Frequency, power and DC current (at DC power
parameters
port) (numerical display only)
Data format
Frequency: Hz, dHz, %, ppm, Hz/V
RF Power: dBm
DC Current: A
Measurement
Set to continuous, hold, or single with internal,
trigger
external, manual, or bus trigger
1
10
100
1k
100 k
10.9
2.7
0.34
0.04
1 M
10.7
2.7
0.34
0.04
10 M
11.1
2.8
0.35
0.04
40 M
17.7
4.4
0.56
0.07
Stop frequency (Hz)
Start frequency (Hz)
1. Measurement time (sec) = 0.2 + the above value x number of correlation when applying
cross-correlation function (standard ONLY). For E5052-011, number of correlation = 1.