System performance with the e5053a downconverter – Atec Agilent-E5052A User Manual

Page 13

background image

13

System Performance with the E5053A Downconverter

Table 2-2. System phase noise sensitivity (typical)

Offset from carrier (Hz)

Input frequency

1

10

100

1k

10 k

100 k

1 M

10 M

40 M

3

GHz

-39.0 -78.7 -99.5 -124.4 -135.3 -137.5 -153.5 -164.5 -167.6

10

GHz

-34.6 -71.8 -91.1 -116.1 -124.1 -128.1 -147.1 -156.6 -160.0

18

GHz

-31.6 -66.5 -85.3 -110.3 -121.2 -125.3 -141.3 -150.7 -154.2

26.5

GHz

-29.0 -63.2 -81.8 -106.8 -117.8 -121.8 -137.8 -147.3 -150.8

E5052A with E5053A phase noise performance (standard,
correlation = 1, measuring ultra-low noise N5507A uW LO)

Block diagram of SSA millimeter wave

Figure 2-2. System phase noise sensitivity with the E5053A downconverter
(standard, 0 dBm input, correlation = 1, start offset frequency = 1 Hz, measurement time = 17.7 sec)