Internal real-time complex digital I/Q filters (included with Option 653 )
Factory channel correction (256 taps)
Corrects the linear phase and amplitude response of the baseband IQ and RF outputs of the signal generator using factory calibration
arrays (default mode is off).
RF amplitude flatness (120 MHz)
± 0.2 dB measured
RF phase flatness (120 MHz)
± 2 degrees measured
User channel correction (256 taps)
Automated routine uses USB power sensor to correct for linear phase and amplitude response of DUT (equalizer). See User Guide for
more details.
Max RF amplitude flatness correction
± 15 dB
Max RF phase flatness correction
± 20 degrees
Equalization filter (256 taps)
User can download and apply inverse or custom phase and amplitude response coefficients from tools such as MATLAB, 89600 VSA,
or SystemVue to correct for linear errors of DUT/system. See User Guide for more details.
Baseband generator (Options 653 and 655)
Channels
2 [I and Q]
Resolution
16 bits [1/65,536]
Sample rate
Option 653
100 Sa/s to 75 MSa/s
Option 653 and 655
100 Sa/s to 150 MSa/s
RF (I+Q) bandwidth
Option 653
60 MHz, nominal
Option 653 and 655
120 MHz, nominal
Interpolated DAC rate
800 MHz (waveforms only need OSR = 1.25 )
Frequency offset range
± 60 MHz
Digital sweep modes
In list sweep mode each point in the list can have independent waveforms (N5172B)
along with user definable frequencies and amplitudes; see the Amplitude and Frequency
Specifications sections for more detail.
Waveform switching speed
1
SCPI mode
≤ 5 ms, measured (standard)
≤ 1.2 ms, measured (Option UNZ)
List/step sweep mode
≤ 5 ms, measured (standard)
≤ 900 us, measured (Option UNZ)
Waveform transfer rates
(measured, no markers)
FTP LAN to internal SSD
10.7 MB/sec or 2.67 Msa/sec
Internal SSD to FTP LAN
7.7 MB/sec 1.92 Msa/sec
FTP LAN to BBG
8.2 MB/sec or 2.05 Msa/sec
FTP LAN to BBG encrypted
4 MB/sec or 1 Msa/sec
USB to BBG
19 MB/sec or 4.75 Msa/sec
BBG to USB
1.2 MB/sec or 300 Ksa/sec
Internal SSD to BBG
48 MB/sec or 12 Msa/sec
BBG to internal SSD
1.2 MB/sec or 300 Ksa/sec
SD card to BBG (Option 006)
BBG to SD card (Option 006)
845 KB/sec or 211 Ksa/sec
1. SCPI mode switching speed applies when waveforms are pre-loaded in list sweep and sample rate ≥ 10 MSa/s.