Atec Agilent-N5171B User Manual
Page 12
12
Residual FM (CW mode, 300 Hz to 3 kHz BW, CCITT, rms)
5 MHz to 6 GHz
< N x 2 Hz (measured) (see N value in frequency band table)
Harmonics (CW mode)
Range
Standard < +4 dBm
Option 1EA < +12 dBm
9 kHz to 3 GHz
< –35 dBc
< –30 dBc
> 3 to 4 GHz
< –35 dBc, typical
< –35 dBc, typical
> 4 to 6 GHz
< –53 dBc, typical
< –40 dBc, typical
Nonharmonics (CW mode)
Range
> 10 KHz offset
Standard (dBc)
9 kHz to < 5 MHz
–65, nominal
5 to < 250 MHz
–75
250 to < 750 MHz
–75
750 MHz to < 1.5 GHz
–72
1.5 to < 3.0 GHz
–66
3 to 6 GHz
–60
Subharmonics (CW mode)
9 kHz to 1.5 GHz
None
> 1.5 to 3 GHz
–77 dBc
> 3 to 6 GHz
–74 dBc
Jitter
1
Carrier frequency
SONET/SDH data rate rms jitter BW
μUI rms, measured Seconds, typical
155 MHz
155 MB/s
100 Hz to 1.5 MHz
140
0.9 ps
622 MHz
622 MB/s
1 KHz to 5 MHz
67
0.11 ps
2.488 GHz
2488 MB/s
5 kHz to 20 MHz
271
0.11 ps
Phase coherence (Option 012)
LO input frequency range
250 MHz to 6 GHz, nominal
LO input power range
0 to +12 dBm, nominal
LO output frequency range
250 MHz to 6 GHz, nominal
LO output power range
0 to +12 dBm, nominal
1. Calculated from phase noise performance in CW mode at +10 dBm. For other frequencies, data rates, or bandwidths, please consult your sales
representative.