Atec Agilent-N5173B User Manual

Page 7

background image

7

SWR (measured CW mode)

Frequency

Attenuator state

0 dB

5 dB and greater

≤ 2 GHz

< 1.7:1

< 1.2:1

> 2 to 8 GHz

< 1.4:1

< 1.4:1

> 8 to 13 GHz

< 1.6:1

< 1.5:1

> 13 to 20 GHz

< 1.8:1

< 1.7:1

> 20 to 40 GHz

< 1.6:1

< 1.4:1

External detector leveling

1

Range

–0.2 mV to –0.5 V (nom)

Bandwidth

10 kHz (typ)

Amplitude switching speed

2

SCPI mode

≤ 2 ms (typ)

Power search SCPI mode

3

< 12 ms (meas)

List/step sweep mode

≤ 2 ms (typ)

User flatness correction

Number of points

3201

Number of tables

Dependent on available free memory in instrument; 10,000 maximum

Entry modes

USB/LAN direct power meter control, LAN to GPIB and USB to GPIB, remote bus, and manual

USB/GPIB power meter control

Sweep modes

See Frequency Specifications section for more detail

1. Not intended for pulsed operation.
2. Time from receipt of SCPI command or trigger signal to amplitude settled within 0.2 dB. Specification does not apply when switching to or from fre-

quencies < 5 MHz, or when ALC level is < 0 dBm, or when frequency crosses 0.002, 0.02, 0.1, 2.0, 3.2, 5.0, 6.4, 8, 10, 12.8, 16, 20, 25.6, or 32 GHz.

3. When ALC is off and power search mode is disabled amplitude switching is < 250 us (meas).