Atec Agilent-N5173B User Manual
Page 7
![background image](/manuals/318651/7/background.png)
7
SWR (measured CW mode)
Frequency
Attenuator state
0 dB
5 dB and greater
≤ 2 GHz
< 1.7:1
< 1.2:1
> 2 to 8 GHz
< 1.4:1
< 1.4:1
> 8 to 13 GHz
< 1.6:1
< 1.5:1
> 13 to 20 GHz
< 1.8:1
< 1.7:1
> 20 to 40 GHz
< 1.6:1
< 1.4:1
External detector leveling
1
Range
–0.2 mV to –0.5 V (nom)
Bandwidth
10 kHz (typ)
Amplitude switching speed
2
SCPI mode
≤ 2 ms (typ)
Power search SCPI mode
3
< 12 ms (meas)
List/step sweep mode
≤ 2 ms (typ)
User flatness correction
Number of points
3201
Number of tables
Dependent on available free memory in instrument; 10,000 maximum
Entry modes
USB/LAN direct power meter control, LAN to GPIB and USB to GPIB, remote bus, and manual
USB/GPIB power meter control
Sweep modes
See Frequency Specifications section for more detail
1. Not intended for pulsed operation.
2. Time from receipt of SCPI command or trigger signal to amplitude settled within 0.2 dB. Specification does not apply when switching to or from fre-
quencies < 5 MHz, or when ALC level is < 0 dBm, or when frequency crosses 0.002, 0.02, 0.1, 2.0, 3.2, 5.0, 6.4, 8, 10, 12.8, 16, 20, 25.6, or 32 GHz.
3. When ALC is off and power search mode is disabled amplitude switching is < 250 us (meas).