Frequency reference

Accuracy

± aging rate

± temperature effects

± line voltage effects

± initial setting accuracy

Internal time base reference oscillator aging rate

1

< ± 1 x 10^-7/year (nom)

< ± 5 x 10^-10/day after 30 days (nom)

Initial achievable calibration accuracy

± 4 x 10^-8 or ± 40 ppb

Adjustment resolution

< 1 x 10^-10 (nom)

Temperature effects

< ± 2 x 10^-8 from 20 to 30 °C (nom)

Line voltage effects

< ± 1 x 10^-9 for ± 10% change (nom)

Reference output
Frequency

10 MHz

Amplitude

≥ +4 dBm, (nom) into 50 Ω load

External reference input
Input frequency standard

10 MHz

Input frequency Option 1ER

1 to 50 MHz (in multiples of 0.1 Hz)

Lock range

± 1 ppm (nom)

Amplitude

5 dBm ± 2 dB (nom)

2

Impedance

50 Ω (nom)

Waveform

Sine or square

Sweep modes (frequency and amplitude)

Operating modes

Step sweep (equally spaced frequency and amplitude or

logarithmically spaced frequency steps)

List sweep (arbitrary list of frequency and amplitude steps)

Simultaneously sweep waveforms with N5182B; see baseband

generator section for more detail

Sweep range

Within instrument frequency range

Dwell time

100 µs to 100s

Number of points

2 to 65535 (step sweep)

1 to 3201 (list sweep)

Step change

Linear or logarithmic

Triggering

Free run, trigger key, external, timer, bus (GPIB, LAN, USB)

1. Aging rate is determined by design as a function of the OCXO.
2. Inputs between +3 dBm to +20 dBm are allowed.