Absolute level accuracy in cw mode, Alc on) () = typical – Atec Agilent-N5173B User Manual
Page 6
![background image](/manuals/318651/6/background.png)
6
-1
-0.8
-0.6
-0.4
-0.2
0
0.2
0.4
0.6
0.8
1
0
2
4
6
8
10
12
14
16
18
20
mean
mean - std dev
mean + std dev
Frequency (GHz) vs. level accuracy (dB)
Measured level accuracy at -110 dBm
Frequency (GHz) vs. level accuracy (dB)
Measured level accuracy at -90 dBm
-1.00
-0.80
-0.60
-0.40
-0.20
0.00
0.20
0.40
0.60
0.80
1.00
0
5
10
15
20
25
30
35
40
mean
mean - std dev
mean + std dev
Absolute level accuracy in CW mode
1, 2
(ALC on) () = typical
With or without Option 1E1
With Option 1E1
Max power to
+10 dBm
< +10 to
–10 dBm
< –10 to
– 20dBm
< –20 to
–75 dBm
< –75 to
–90 dBm
< –90 to
–120 dBm
9 kHz to 2 GHz
± 0.6 dB
± 0.6 dB
± 0.7 dB
± 0.7 dB
± 1.4 dB
(± 0.3)
> 2 to 20 GHz
± 0.9 dB
± 0.7 dB
± 0.7 dB
± 0.7 dB
± 1.6 dB
(± 0.3)
> 20 to 40 GHz
± 0.9 dB
± 0.8 dB
± 1.1 dB
± 1.1 dB
± 2.0 dB
1. Level accuracy applies between 15 °C and 35 °C. Specifications do not apply above the maximum specified power. For temperatures outside this
range, absolute level accuracy degrades by 0.01 dB/degree C for frequencies ≤ 4.5 GHz and 0.02 dB/degree C for frequencies > 4.5 GHz.
2. For instruments with Type-N connectors (Option 1ED), specifications are degraded typically 0.2 dB above 18 GHz.
-1
-0.8
-0.6
-0.4
-0.2
0
0.2
0.4
0.6
0.8
1
0
2
4
6
8
10
12
14
16
18
20
mean
mean - std dev
mean + std dev
Frequency (GHz) vs. level accuracy (dB)
Measured level accuracy at -120 dBm