beautypg.com

Absolute level accuracy in cw mode, Alc on) () = typical – Atec Agilent-N5173B User Manual

Page 6

background image

6

-1

-0.8

-0.6

-0.4

-0.2

0

0.2

0.4

0.6

0.8

1

0

2

4

6

8

10

12

14

16

18

20

mean

mean - std dev

mean + std dev

Frequency (GHz) vs. level accuracy (dB)

Measured level accuracy at -110 dBm

Frequency (GHz) vs. level accuracy (dB)

Measured level accuracy at -90 dBm

-1.00

-0.80

-0.60

-0.40

-0.20

0.00

0.20

0.40

0.60

0.80

1.00

0

5

10

15

20

25

30

35

40

mean

mean - std dev

mean + std dev

Absolute level accuracy in CW mode

1, 2

(ALC on) () = typical

With or without Option 1E1

With Option 1E1

Max power to

+10 dBm

< +10 to

–10 dBm

< –10 to

– 20dBm

< –20 to

–75 dBm

< –75 to

–90 dBm

< –90 to

–120 dBm

9 kHz to 2 GHz

± 0.6 dB

± 0.6 dB

± 0.7 dB

± 0.7 dB

± 1.4 dB

(± 0.3)

> 2 to 20 GHz

± 0.9 dB

± 0.7 dB

± 0.7 dB

± 0.7 dB

± 1.6 dB

(± 0.3)

> 20 to 40 GHz

± 0.9 dB

± 0.8 dB

± 1.1 dB

± 1.1 dB

± 2.0 dB

1. Level accuracy applies between 15 °C and 35 °C. Specifications do not apply above the maximum specified power. For temperatures outside this

range, absolute level accuracy degrades by 0.01 dB/degree C for frequencies ≤ 4.5 GHz and 0.02 dB/degree C for frequencies > 4.5 GHz.

2. For instruments with Type-N connectors (Option 1ED), specifications are degraded typically 0.2 dB above 18 GHz.

-1

-0.8

-0.6

-0.4

-0.2

0

0.2

0.4

0.6

0.8

1

0

2

4

6

8

10

12

14

16

18

20

mean

mean - std dev

mean + std dev

Frequency (GHz) vs. level accuracy (dB)

Measured level accuracy at -120 dBm