background image

10 MHz OCXO fitted as standard

Ageing Rate

< ± 0.8 x 10

-7

per year after 30 days continuous use

Temperature Coefficient

< ± 5 x 10

-8

over the temperature range 0°C to 50°C

Output Frequency

Within 2 x 10

-7

of final frequency after 10 minutes from connecting

supply power and switching on at a temperature of 20°C

Standby power is provided while the instrument is off but connected to
the supply.

Output of 2 V pk-pk from 50

is provided on a rear panel BNC con-

nector.

1 MHz or 10 MHz at a level of 300 mV RMS to 1.8 V RMS into
1 k

on the rear panel BNC connector

With this option fitted RF output, EXT I/EXT AM input and EXT Q/EXT FM
input connectors are transferred to the rear panel. When Option 009
is fitted only RF output connector is transfered to the rear panel. The
standard signal generator specification remains unaltered.

WARRANTY

2 years with options for 3, 4 and 5 years

CALIBRATION INTERVAL

Recommended at 2 years

REMOTE CONTROL INTERFACES

Ethernet

All signal generator parameters except the supply switch are remotely
programmable. The following LAN protocols supported:

VXI-11
Telephone Network (TELNET)
File Transfer Protocol (FTP) (instrument firmware upgrades only)

GPIB

All signal generator parameters except the supply switch are remotely
programmable. The GPIB is designed in accordance with the IEEE
488.2.

RS-232

All functions except the supply switch are remotely programmable.
Can be used for upgrading the instrument firmware without removal of
the instrument covers.

Interface Functions

SH1, AH1, T6, L4, SR1, RL1, PP0, DC1, DT1, C0, E2

DIMENSIONS AND WEIGHT

Height

Width

Depth

Overall

107 mm

468 mm (19")

545 mm max

Rackmount * 89 mm

425 mm

545 mm max

* Occupies 2U of rack height excluding removable feet bottom feet and front handles

Weight

3412/13/14

10.5 kg

3416

11.5 kg

RATED RANGE OF USE

MIL-T-28800E Class 5

Temperature

0°C to 50°C

Humidity

45%, 0°C to 50°C
95%, 30°C to 40°C

Altitude

700 mbars (3050 m, 10,000 feet)

CONDITIONS OF STORAGE AND TRANSPORT

MIL-T-28800E Class 5
Temperature -40°C to + 71°C
Altitude 570 mbar (4570 m, 15,000 feet)

POWER REQUIREMENTS

AC Supply

100 - 240 V ~ (Limit 90 - 264 V)
50 - 60 Hz ~ (Limit 45 -66 Hz)
185 VA Max

ELECTROMAGNETIC COMPATIBILITY

Conforms to EC directives 89/336/EEC and standard IEC/EN 61326-
1:1997;RF emission class B, immunity table 1 and performance crite-
rion B

SAFETY

Conforms with the requirements of EEC Council Directive 73/23/EEC
(as amended) and the product safety standard IEC / EN61010-1 :
2001 + C1 : 2002 + C2 : 2003 for class 1 portable equipment, for
use in a Pollution Degree 2 environment. The instrument is designed
to be operated from an Installation Category 2 supply.

3412

250 kHz to 2 GHz Digital RF Signal Generator

3413

250 kHz to 3 GHz Digital RF Signal Generator

3414

250 kHz to 4 GHz Digital RF Signal Generator

3416

250 kHz to 6 GHz Digital RF Signal Generator

Supplied with AC power supply lead and CD-ROM
containing:

Operating Manual

Data Sheet

Factor y Test Results (for the unit supplied) and
Certificate of Calibration

ARB data file creation and download soft-

ware

VISA Plug 'n' Play driver software

Librar y of common data files for dual ARB option