Atec Aeroflex-IFR-3416 User Manual
Page 11
![background image](/manuals/318646/11/background.png)
10 MHz OCXO fitted as standard
Ageing Rate
< ± 0.8 x 10
-7
per year after 30 days continuous use
Temperature Coefficient
< ± 5 x 10
-8
over the temperature range 0°C to 50°C
Output Frequency
Within 2 x 10
-7
of final frequency after 10 minutes from connecting
supply power and switching on at a temperature of 20°C
Standby power is provided while the instrument is off but connected to
the supply.
Output of 2 V pk-pk from 50
is provided on a rear panel BNC con-
nector.
1 MHz or 10 MHz at a level of 300 mV RMS to 1.8 V RMS into
1 k
on the rear panel BNC connector
With this option fitted RF output, EXT I/EXT AM input and EXT Q/EXT FM
input connectors are transferred to the rear panel. When Option 009
is fitted only RF output connector is transfered to the rear panel. The
standard signal generator specification remains unaltered.
WARRANTY
2 years with options for 3, 4 and 5 years
CALIBRATION INTERVAL
Recommended at 2 years
REMOTE CONTROL INTERFACES
Ethernet
All signal generator parameters except the supply switch are remotely
programmable. The following LAN protocols supported:
VXI-11
Telephone Network (TELNET)
File Transfer Protocol (FTP) (instrument firmware upgrades only)
GPIB
All signal generator parameters except the supply switch are remotely
programmable. The GPIB is designed in accordance with the IEEE
488.2.
RS-232
All functions except the supply switch are remotely programmable.
Can be used for upgrading the instrument firmware without removal of
the instrument covers.
Interface Functions
SH1, AH1, T6, L4, SR1, RL1, PP0, DC1, DT1, C0, E2
DIMENSIONS AND WEIGHT
Height
Width
Depth
Overall
107 mm
468 mm (19")
545 mm max
Rackmount * 89 mm
425 mm
545 mm max
* Occupies 2U of rack height excluding removable feet bottom feet and front handles
Weight
3412/13/14
10.5 kg
3416
11.5 kg
RATED RANGE OF USE
MIL-T-28800E Class 5
Temperature
0°C to 50°C
Humidity
45%, 0°C to 50°C
95%, 30°C to 40°C
Altitude
700 mbars (3050 m, 10,000 feet)
CONDITIONS OF STORAGE AND TRANSPORT
MIL-T-28800E Class 5
Temperature -40°C to + 71°C
Altitude 570 mbar (4570 m, 15,000 feet)
POWER REQUIREMENTS
AC Supply
100 - 240 V ~ (Limit 90 - 264 V)
50 - 60 Hz ~ (Limit 45 -66 Hz)
185 VA Max
ELECTROMAGNETIC COMPATIBILITY
Conforms to EC directives 89/336/EEC and standard IEC/EN 61326-
1:1997;RF emission class B, immunity table 1 and performance crite-
rion B
SAFETY
Conforms with the requirements of EEC Council Directive 73/23/EEC
(as amended) and the product safety standard IEC / EN61010-1 :
2001 + C1 : 2002 + C2 : 2003 for class 1 portable equipment, for
use in a Pollution Degree 2 environment. The instrument is designed
to be operated from an Installation Category 2 supply.
3412
250 kHz to 2 GHz Digital RF Signal Generator
3413
250 kHz to 3 GHz Digital RF Signal Generator
3414
250 kHz to 4 GHz Digital RF Signal Generator
3416
250 kHz to 6 GHz Digital RF Signal Generator
Supplied with AC power supply lead and CD-ROM
containing:
Operating Manual
Data Sheet
Factor y Test Results (for the unit supplied) and
Certificate of Calibration
ARB data file creation and download soft-
ware
VISA Plug 'n' Play driver software
Librar y of common data files for dual ARB option