beautypg.com

Service, Principles of operation, Troubleshooting – Atec Agilent-ECP-E26A User Manual

Page 36

background image

1

Introduction

20

Agilent E4412A and E4413A Power Sensors Operating and Service Guide

Service

Service instructions consist of principles of operation, troubleshooting, and
repairs.

Principles of Operation

The A1 Bulkhead assembly provides a 50 ohm load to the RF signal applied to
the power sensor. A diode assembly in the bulkhead rectifies the applied RF to
produce a dc voltage which varies with the RF power across the 50 ohm load.
Thus the voltage varies with the RF power dissipated in the load. With
maximum specified RF power (100 mW) the dc voltage is approximately 1V.

The low- level dc voltage from the bulkhead assembly must be amplified before
it can be transferred on standard cables to the power meter. The amplification
is provided by an input amplifier assembly which consists of a chopper
(sampling gate) and an input amplifier. The chopper circuit converts the dc
voltage to an ac voltage. To do this, the chopper uses two field effect
transistors (FETs), A2Q1 and A2Q2, controlled by a 440 Hz square wave
generated by the power meter. The amplitude of the sampling gate output
(drain of A2Q1, source of A2Q2) is a 440 Hz square wave which varies with the
RF power input. The 440 Hz ac output is applied to the input amplifier A2Q3
which provides the input to the first amplifier stage in the power meter.

The E44XX- Series power meter automatically detects when an E44XX Series
power sensor is connected and downloads the correction data from the
sensor’s EEPROM. This configures the power meter to operate over the +20

dBm to

-

70 dBm power range with that particular sensor’s unique

correction

data applied.

Troubleshooting

Troubleshooting information is intended to first isolate the power sensor, the
cable, or the power meter as the defective component. When the power sensor
is isolated, a “Restored Sensor Module” must be used for repair. See

Table 6

.