11

Installation

Initial Inspection

Inspect the shipping container for damage. If the shipping container or
packaging material is damaged, it should be kept until the contents of the
shipment have been checked mechanically and electrically. If there is
mechanical damage or if the instrument does not pass the performance tests,
notify the nearest Agilent Technologies office. Keep the damaged shipping
materials (if any) for inspection by the carrier and a Agilent Technologies
representative.

Interconnections

Connect one end of the 11730A sensor cable to the E4412A or E4413A power
sensor and co nnect the other end of the cable to the power meter’s channel
input. Allow a few seconds for the power meter to download the power
sensor’s calibration table before making a measurement.

Measurement connector (connects to DUT)

E4412A: Type- N (male)

E4413A: 3.5- mm (male)

A torque wrench should be used to tighten these connectors. Use a 3/4- inch
open- end wrench and torque to 12 in- lb (135 Ncm) for the Type- N connector
Use a 20- mm open- end wrench and torque to 8 in- lb (90 Ncm) for the 3.5- mm
connector.

Recommended Calibration Interval

Agilent Technologies recommends a one- year calibration cycle for the E4412A
and E4413A power sensors.