Performance test, Table 4 – Atec Agilent-ECP-E26A User Manual
Page 31
Introduction
1
Agilent E4412A and E4413A Power Sensors Operating and Service Guide
15
Performance Test
Standing Wave Ratio (SWR) and Reflection Coefficient
(Rho)Performance Test
This section does not establish preset SWR test procedures since there are
several test methods and different equipm ent available for testing the SWR or
reflection coefficient. Therefore, the actual accuracy of the test equipment
must be accounted for when measuring against instrument specifications to
determine a pass or fail condition. The test system used must not exceed the
system Rho uncertainties shown in
when testing the E4412A, or in
when testing the E4413A
Table 4
Power Sensor SWR and Reflection Coefficient for the E4412A
a
a.Specifications apply to instruments with serial prefix US3848/MYxxxxxxxx/SGxxxxxxxx and
above. For earlier instruments, refer to Appendix A.
Frequency
System Rho
Uncertainty
Actual
Measurement
Maximum Rho
10 MHz to < 30 MHz
± 0.010
0.099
30 MHz to < 2 GHz
± 0.010
0.070
2 GHz to < 6 MHz
± 0.010
0.078
6 GHz to < 11 GHz
± 0.010
0.091
11 GHz to 18 GHZ
± 0.010
0.119