beautypg.com

Infiniium performance characteristics, Continued) – Atec Agilent-DSO8104A User Manual

Page 23

background image

23

Infiniium Performance Characteristics

(continued)

General characteristics

DSO8064A, MSO8064A, DSO8104A, MSO8104A

Temperature

Operating

0°C to + 50°C

Non-operating

–40°C to + 70°C

Humidity

Operating

Up to 95% relative humidity (non-condensing) at +40°C

Non-operating

Up to 90% relative humidity at +65°C

Altitude

Operating

Up to 4,600 meters (15,000 feet)

Non-operating

Up to 15,300 meters (50,000 feet)

Vibration

Operating

Random vibration 5-500 Hz, 10 minutes per axis, 0.3 g(rms)

Non-operating

Random vibration 5-500 Hz, 10 minutes per axis, 2.41 g(rms); resonant search 5-500 Hz,
swept sine, 1 octave/minute sweep rate, (0.75g), 5 minute resonant dwell at 4 resonances
per axis

Power

100-240 VAC, ± 10%, Cat II, 47 to 440 Hz; Max power dissipated: 390 W

Weight

Net: 13.4 kg (29.5 lbs.)
Shipping: 16.4 kg (36.1 lbs.)

Dimensions (excluding handle)

Height: 216 mm (8.5 in); width: 437 mm (17.19 in); depth: 440 mm (17.34 in)

Safety

Meets IEC1010-1 +A2, CSA certified to C22.2 No.1010.1, Self certified to UL 3111

*

Denotes Warranted Specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ±10°C from firmware calibration temperature.

1

Rise time figures are calculated from t r = 0.35/bandwidth.

2

Magnification is used below 5 mV/div range. Below 5 mV/div, full scale is defined as 40 mV. Full scale is defined as the major attenuator setting above an intermediate setting.
(Major settings 50

Ω: 10 mV, 20 mV, 50 mV, 100 mV, 200 mV, 500 mV, 1 V, 1 MΩ: all of the above plus 2 V).

3

Vertical resolution for 8 bits = 0.4% of full scale, for 12 bits = 0.024% of full scale.

4

The dc gain accuracy decreases 0.08% of full scale per degree C from the calibration temperature.

5

Maximum 2-channel memory depth only available at maximum 2-channel sample rate. Maximum each channel memory depth available at any selectable sample rate.

6

Valid for vertical ranges > 5 mV / div.

7

FFT amplitude readings are affected by input amplifier roll-off; DSO8064A and MSO8064A: –3 dB at 600 MHz, with amplitude decreasing as frequency increases above 600 MHz;
DSO8104A and MSO8104A: –3 dB at 1 GHz, with amplitude decreasing as frequency increases above 1 GHz.

8

Noise floor varies with memory depth and with averaging on or off.