Infiniium Performance Characteristics
(continued)
Acquisition: scope channels
DSO8064A, MSO8064A, DSO8104A, MSO8104A
Real time sample rate (max)
2 channels
4 GSa/s
Each channel
2 GSa/s
Equivalent time sample rate (max)
250 GSa/s
Memory depth
2 channels/each channel
Standard
1 M / 500 K
Option 040
8 M / 4 M
Option 080
16 M / 8 M
Option 160
32 M / 16 M
Option 320
64 M / 32 M
Option 640
128 M / 64 M
Sampling modes
Real time
Normal
Successive single-shot acquisitions
Peak detect
Captures and displays narrow pulses or glitches at all real time sample rates
High resolution
Real-time boxcar averaging reduces random noise and increases resolution
Equivalent time
Random repetitive sampling (higher time resolution at faster sweep speeds)
Segmented memory
Captures bursting signals at maximum sample rate without consuming memory during
periods of inactivity. Selectable number of segments up to 32,768 depending on
memory option installed. Minimum inter-segment time (or the time between the end of
the previous acquisition and the beginning of the next acquisition) of 20 µs.
Averaging
Selectable from 2 to 4096
Filters
Sin[x])/x Interpolation
Filter On/Off selectable FIR digital filter. Digital signal processing adds points between
acquired data points to enhance measurement accuracy and waveform display quality.
BW = sample rate/4
Acquisition: digital channels
MSO8064A, MSO8104A
Maximum real time sample rate
1 GSa/s
Memory depth per channel
32 M
Minimum width glitch detection
2.5 ns