beautypg.com

Fieldfox rf vector network analyzer, Description specification supplemental information – Atec Agilent-N9923A User Manual

Page 3

background image

3

FieldFox RF Vector Network Analyzer

Description

Specification

Supplemental Information

Frequency range

2 MHz to 4 GHz

Option 104

2 MHz to 6 GHz

Option 106

Frequency reference

Accuracy

±2 ppm

Aging rate

±1 ppm/yr

Temperature stability

±1 ppm over -10 to 55 ºC

Frequency resolution

1 Hz < 3.2 GHz
1.2 Hz > 3.2 GHz

Data points

101, 201, 401, 601, 801, 1001, 1601, 4001, 10,001 (custom number of
points can be set using SCPI)

Measurements

S11, S21 magnitude and phase
S22, S12 magnitude and phase (option 122)

Formats

Log magnitude, linear magnitude, VSWR,
phase, Smith chart, polar, group delay,
unwrapped phase, real, imaginary

Directivity

Corrected, with full two-port
calibration, N9910X-800 calibration
kit (option 122)

42 dB

Corrected with QuickCal (option 112)

38 dB (typical)

System impedance

50Ω (nominal)

75Ω with appropriate adapter and
calibration kit

System dynamic range

1

(IF bandwidth = 300 Hz, 32 averages)

2 MHz to 6 GHz

90 dB

100 dB (typical)

Receiver dynamic range

2

(IF bandwidth = 300 Hz)

2 MHz to 6 GHz

104 dB (typical)

Receiver compression level at 0.1 dB compression

+10 dBm (typical)

Test port output power

High

+5 dBm (nominal)

Low

-40 dBm (nominal)

Manual

Power settable in 1 dB steps

Test port 1 or 2 damage level

2 MHz to 6 GHz

+23 dBm

Trace noise (high output power, IF bandwidth = 300 Hz)

Magnitude

< 3 GHz, 0.008 dB rms (typical)
> 3 GHz, 0.01 dB rms (typical)

Temperature stability

Stability magnitude

0.01 dB/ºC at 23 ºC ±5 ºC (typical)

IF bandwidths

300 Hz, 1 kHz, 3 kHz, 10 kHz, 30 kHz

1. System dynamic range = source maximum output power - receiver noise floor

2. Receiver dynamic range = receiver maximum input level - receiver noise floor