beautypg.com

General information, Description specification supplemental information – Atec Agilent-N9923A User Manual

Page 14

background image

14

General Information

(continued)

Description

Specification

Supplemental Information

Battery

10.8 V, 4.6 A–h

Lithium ion

Operating time

3.5 hours (typical)

Charge time

A fully discharged battery takes
about 1.5 hours to recharge to
80%, 4 hours to 100%

Discharge temperature limits

1

–10 to 60 ºC, ≤ 85% RH

Charge temperature limits

1

0 to 45 ºC, ≤ 85% RH

Storage temperature limits

–20 to 50 ºC1, ≤ 85% RH

The battery packs should be stored in an
environment with low humidity. Extended
exposure to temperature above 45 ºC could
degrade battery performance and life

Data storage

Internal

Minimum 16 MB

Up to 1000 instrument states and trace

External

Supports USB 2.0 compatible memory
devices; Supports microSD and microSDHC
memory cards

Display

6.5” transflective color VGA
LED–backlit
640 x 480 with anti–glare coating

Weight

2.7 kg (6.0 lbs) including battery

Dimensions (H x W x D)

292 x 188 x 72 mm
(11.5” x 7.4” x 2.8”)

Test ports

RF Port 1 or Port 2

Connector

Type–N, female

Impedance

50 Ω (nominal)

Damage level

> +23 dBm, > ±50 VDC

1. Charge and discharge temperatures are internal temperatures of the battery as measured by a sensor embedded in the

battery. The Battery screen displays temperature information. To access the screen, select System ,

Service Diagnostics

,

and

Battery