Multiplexer selection guide – Atec Agilent-34972A User Manual
Page 21
21
Multiplexer selection guide
Choose between the broad functionality of the 34901A, the
high speed scanning of the 34902A, or the single-ended
density of the 34908A. These three modules are the only
way to connect to the 34970A/34972A internal DMM. They
can be used to scan with external instruments as well.
All multiplexer modules employ break-before-make
scanning, ensuring only one closed channel (or channel
pair) at a time. Multiple channel closures are allowed on
the 34901A and 34902A modules when not configured
for scanning.
The 34908A does not allow multiple channel closures
at any time.
34901A
20-Channel General Purpose Multiplexer
• 60 ch/s scanning
• Two- and four-wire scanning
• Built-in thermocouple reference
junction
• 300 V switching
The Agilent 34901A is the most
versatile multiplexer for general
purpose scanning. It combines
dense, multifunction switching with
60-channel/second scan rates to
address a broad spectrum of data
acquisition applications.
Two- and four-wire channels can
be mixed on the same module. Two
additional fused inputs (22 channels
total) route up to 1 A of current to
the internal DMM, allowing ac and dc
current measurements without the
need for external shunt resistors.
34901A
34902A
34908A
Number of channels
20 + 2
16
40
Max scan speed
60 ch/s
250 ch/s
60 ch/s
Number of contacts
2 or 4
2 or 4
1
Temperature
Thermocouple
●
●
●
2-wire RTD
●
●
●
4-wire RTD
●
●
Thermistor
●
●
●
dc Volts
●
●
●
ac Volts
●
●
●
2-wire Ohms
●
●
●
4-wire Ohms
●
●
Frequency
●
●
●
Period
●
●
●
dc current
●
ac current
●
Backplane switches
Internal
DMM input
01
10
Com
H
L
H
L
H
L
H
L
H
L
H
L
I
L
I
L
I
L
I
L
H
L
H
L
T
Com
(4W sense)
11
20
21
Current
channels
22
Com
(Current)
Fuse
Fuse
Internal
DMM input
(4W sense)
Internal
DMM input
(Current)
Reference
junction
sensor
Channel switches
Bank switch
Shunt switches