beautypg.com

Measurement characteristics, Operating characteristics – Atec Agilent-34972A User Manual

Page 17

background image

17

Measurement characteristics

[7]

DC voltage
Measurement Method

Continuously Integrating

Multi-slope III A-D converter

A/D linearity

0.0002% of reading + 0.0001 % of range

Input resistance
100 mV, 1 V, 10 V ranges Selectable 10 MΩ or > 10,000 MΩ
100 V, 300 V ranges

10 MΩ ± 1%

Input bias current

< 30 pA at 25 °C

Input protection

300 V all ranges

True RMS AC voltage
Measurement method

AC coupled True RMS — measures the

AC component of the input with up to

300 Vdc of bias on any range

Crest factor

Maximum of 5:1 at Full Scale

Additional crest factor
errors (non-sinewave)

Crest Factor 1-2

0.05 % of reading

Crest Factor 2-3

0.15 % of reading

Crest Factor 3-4

0.30 % of reading

Crest Factor 4-5

0.40 % of reading

Input impedance

1 MΩ ± 2% in parallel with 150 pF

Input protection

300 Vrms all ranges

Resistance
Measurement method

Selectable 4-wire or 2-wire Ohms

Current source referenced to LO input

Offset compensation

Selectable on 100 Ω, 1 kΩ, 10 kΩ ranges

Maximum lead resistance

10% of range per lead for 100 Ω and

1 kΩ ranges. 1 kΩ on all other ranges

Input Protection

300 V on all ranges

Frequency and period
Measurement method

Reciprocal counting technique

Voltage ranges

Same as AC Voltage function

Gate time

1s, 100 ms, or 10 ms

Measurement timeout

Selectable 3 Hz, 20 Hz, 200 Hz LF limit

DC Current
Shunt resistance

5Ω for 10 mA, 100 mA; 0.1 Ω for 1 A

Input protection

1A 250 V fuse on 34901A module

True RMS AC current
Measurement method

Direct coupled to the fuse and shunt.

AC coupled True RMS measurement

(measures the ac component only)

Shunt resistance

5 Ω for 10 mA; 0.1 Ω for 100 mA, 1 A

Input protection

1 A 250 V fuse on 34901A module

Thermocouple
Conversion ITS-90

software

compensation

Reference junction type

Internal, Fixed, or External

Open thermocouple check

Selectable per channel. Open > 5 kΩ

Thermistor

44004, 44007, 44006 series

RTD

α = 0.00385 (DIN) and α = 0.00391

Measurement noise rejection 60 (50) Hz

[1]

dc CMRR

140 dB

ac CMRR

70 dB

Integration time

Normal mode rejection

[2]

200 plc/3.33s (4s)

110 dB

[3]

100 plc/1.67s (2s)

105 dB

[3]

20 plc/333 ms (400 ms)

100 dB

[3]

10 plc/167 ms (200 ms)

95 dB

[3]

2 plc/33.3 ms (40 ms)

90 dB

1 plc/16.7 ms (20 ms)

60 dB

< 1 plc

0 dB

Operating characteristics

[4]

Single channel measurement rates

[5]

Function

Resolution

[8]

34970A/34972A

readings/sec

dcV, 2-wire resistance

6½ digits (10 plc)

6 (5)

5½ digits (1 plc)

54 (47)

4½ digits (0.02 plc)

500

Thermocouple

0.1 °C (10 plc)

6 (5)

0.1 °C (1 plc)

52 (47)

(0.02 plc)

280

RTD, thermistor

0.01 °C (10 plc)

6 (5)

0.1 °C (1 plc)

49 (47)

1 °C (0.02 plc)

200

acV

6½ Slow (3 Hz)

0.14

6½ Med (20 Hz)

1

6½ Fast (200 Hz)

8

[6]

100

Frequency, period

6½ digits (1 s gate)

1

5½ digits (100 ms)

9

4½ digits (10 ms)

70

[1]

For 1 KΩ unbalance in LO lead

[2]

For power line frequency ± 0.1%

[3]

For power line frequency ± 1% use 80 dB or ± 3% use 60 dB

[4]

Reading speeds for 60 Hz and (50 Hz) operation

[5] For

fi xed function and range, readings to memory, scaling and

alarms off, AZERO OFF, USB datalogging OFF

[6]

Maximum limit with de fault settling delays defeated

[7]

Isolation voltage (ch-ch, ch-earth) 300 Vdc, ac rms

[8]

6½ digits = 22 bits, 5½ digits = 18 bits, 4½ digits = 15 bits