Single channel reading rates to I/O or internal memory
(See footnotes for conditions below)
34970A
34972A
into memory
readings/sec
to GPIB or RS232
readings/sec
to LAN, USB or
memory
readings/sec
Single channel ASCII readings
500
440
500
Single channel while changing scale
(eg MEAS dcV 10/MEAS dcV 1)
25
25
25
Single channel while changing function
(eg MEAS dcV/MEAS Ohms)
12
12
12
Scanning measurement rates to I/O or internal memory
(See footnotes for conditions below)
34970A
34972A
into memory
ch/sec
to GPIB or RS232
ch/sec
to LAN, USB or
memory
ch/sec
Scanning dcV or ohms channels
34901A/34908A
60
60
60
34902A
250
210
240
34902A into and out of memory (using INIT, FETCh)
—
180
240
34902A with timestamp (using MEAS)
—
150
240
34902A with scaling and alarms
220
190
220
34902A dcV and ohms on alternate channels
80
80
80
Scanning acV channels
(2)
34901A/34908A
50
50
50
34902A
100
90
100
Scanning temperature – thermistor or TC channels
34901A/34908A
50
50
50
34902A
160
150
150
Scanning digital in/totalizer channels
34907A Digital Input
275
250
275
34907A Totalizer
240
210
240
Data out of memory
(3)(4)
(FETCh of 50K readings)
(See footnotes for conditions below)
34970A
34972A
Single channel
[1] [2]
over GPIB
readings/sec
over RS232
readings/sec
over USB
readings/sec
over LAN or
memory
readings/sec
Readings
800
600
55 K
120 K
Readings with timestamp
450
320
35 K
60 K
Readings with all format options ON
310
230
25 K
50 K
[1]
Speeds are for 4 1/2 digits, delay 0, display off, USB data logging off, autozero off, auto-ranging off unless otherwise noted. Use MEAS command
for best I/O performance. RS232 at 115 Kbaud.
[2]
Maximum, with default delays defeated
[3]
Assumes relative time format (time since start of scan)
[4]
Typical rates assuming lightly loaded PC and limited other traffi c on I/Os. LAN rates assume use of socket connection; VXI11 will be less.
[5] For
fi xed function and range, readings to memory, scaling/alarms/autozero off