Single channel reading rates to I/O or internal memory

(See footnotes for conditions below)

34970A

34972A

into memory

readings/sec

to GPIB or RS232

readings/sec

to LAN, USB or

memory

readings/sec

Single channel ASCII readings

500

440

500

Single channel while changing scale
(eg MEAS dcV 10/MEAS dcV 1)

25

25

25

Single channel while changing function
(eg MEAS dcV/MEAS Ohms)

12

12

12

Scanning measurement rates to I/O or internal memory

(See footnotes for conditions below)

34970A

34972A

into memory

ch/sec

to GPIB or RS232

ch/sec

to LAN, USB or

memory

ch/sec

Scanning dcV or ohms channels
34901A/34908A

60

60

60

34902A

250

210

240

34902A into and out of memory (using INIT, FETCh)

180

240

34902A with timestamp (using MEAS)

150

240

34902A with scaling and alarms

220

190

220

34902A dcV and ohms on alternate channels

80

80

80

Scanning acV channels

(2)

34901A/34908A

50

50

50

34902A

100

90

100

Scanning temperature – thermistor or TC channels
34901A/34908A

50

50

50

34902A

160

150

150

Scanning digital in/totalizer channels
34907A Digital Input

275

250

275

34907A Totalizer

240

210

240

Data out of memory

(3)(4)

(FETCh of 50K readings)

(See footnotes for conditions below)

34970A

34972A

Single channel

[1] [2]

over GPIB

readings/sec

over RS232

readings/sec

over USB

readings/sec

over LAN or

memory

readings/sec

Readings

800

600

55 K

120 K

Readings with timestamp

450

320

35 K

60 K

Readings with all format options ON

310

230

25 K

50 K

[1]

Speeds are for 4 1/2 digits, delay 0, display off, USB data logging off, autozero off, auto-ranging off unless otherwise noted. Use MEAS command
for best I/O performance. RS232 at 115 Kbaud.

[2]

Maximum, with default delays defeated

[3]

Assumes relative time format (time since start of scan)

[4]

Typical rates assuming lightly loaded PC and limited other traffi c on I/Os. LAN rates assume use of socket connection; VXI11 will be less.

[5] For

fi xed function and range, readings to memory, scaling/alarms/autozero off