Modules specifi cations – Atec Agilent-34972A User Manual
Page 20
20
Modules specifi cations
The Agilent 34970A/34972A accuracy specifications already
include the switching offset and reference junction errors
shown below. These errors are listed separately for deter-
mining system error with external measurement devices.
Up to three modules, in any combination, can be inserted
into a single mainframe. The 34970A/34972A’s internal
DMM connections are accessible only through the 34901A,
34902A, and 34908A low-frequency multiplexers.
On-module screw terminals accept wire sizes from
16 gage to 22 gage. Twenty-gage wire is recommended for
high channel count applications. The 34905A and 34906A
RF Multiplexers use SMB connectors. A standard set of
(10) BNC-to-SMB adapter cables is provided with each
RF module for convenient BNC connections.
Multiplexer
Actuator
Matrix
RF multiplexer
Multifunction
34901A
34902A
[1]
34908A
34903A
34904A
34905A
34906A
34907A
General
Number of channels
20 + 2
16
40
20
4 x 8
Dual 1 x 4
See page 25
for module
specifi cations
2/4 wire
2/4 wire
1 wire
SPDT
2 wire
50 Ω
75 Ω
Connects to internal DMM
●
●
●
Scanning speed
60 ch/s
250 ch/s
[1]
60 ch/s
Open/close speed
120/s
120/s
70/s
120/s
120/s
60/s
Input
Voltage (dc , ac rms)
[2]
300 V
300 V
300 V
300 V
300 V
42 V
Current (dc , ac rms)
1 A
50 mA
1 A
1 A
1 A
0.7 A
Power (W , VA)
50 W
2 W
50 W
50 W
50 W
20 W
DC characteristics
Offset voltage
[3]
< 3 uV
< 6 uV
< 3 uV
< 3 uV
< 3 uV
< 6 uV
Initial closed channel R
[3]
< 1 Ω
< 1 Ω
< 1 Ω
< 0.2 Ω
< 1 Ω
< 0.5 Ω
Isolation ch-ch, ch-earth
> 10 GΩ
> 10 GΩ
> 10 GΩ
> 10 GΩ
> 10 GΩ
> 1 GΩ
AC characteristics
Bandwidth
[4]
10 MHz
10 MHz
10 MHz
10 MHz
10 MHz
2 GHz
[5]
2 GHz
[5]
Insertion loss (dB) 10 MHz
—
—
—
—
—
-0.1
-0.1
100 MHz
—
—
—
—
—
-0.4
-0.4
500 MHz
—
—
—
—
—
-0.6
-0.5
1 GHz
—
—
—
—
—
-1
-1
1.5 GHz
—
—
—
—
—
-1.2
-1.5
2 GHz
—
—
—
—
—
-3
-2
SWR 0-9.99 MHz
—
—
—
—
—
1.02
1.02
10-99.99 MHz
—
—
—
—
—
1.05
1.05
100-499.99 MHz
—
—
—
—
—
1.20
1.25
500-999.99 MHz
—
—
—
—
—
1.20
1.40
1-1.499 GHz
—
—
—
—
—
1.30
1.40
1.5-2 GHz
—
—
—
—
—
1.40
2.00
ch-ch cross talk (dB)
[4]
10MHz
-45
-45
-18
[6]
-45
-33
-100
-85
100 MHz
—
—
—
—
—
-85
-75
500 MHz
—
—
—
—
—
-65
-65
1 GHz
—
—
—
—
—
-55
-50
1.5 GHz
—
—
—
—
—
-45
-40
2 GHz
—
—
—
—
—
-35
-35
Risetime
< 300 ps
Signal delay
< 3 ns
Capacitance HI - LO
< 50 pF
< 50 pF
< 50 pF
< 10 pF
< 50 pF
< 20 pF
LO - Earth
< 80 pF
< 80 pF
< 80 pF
< 80 pF
< 80 pF
—
Volt-Hertz limit
10
8
10
8
10
8
10
8
10
8
10
10
Other
T/C cold junction accuracy
[3]
(typical)
0.8 °C
0.8 °C
0.8 °C
[8]
Switch life No load (typical)
100 M
100 M
100 M
100 M
100 M
5 M
5 M
Rated load (typical)
[7]
100 k
100 k
100 k
100 k
100 k
100 k
100 k
Temperature Operating
all cards — 0 °C to 55 °C
Storage
all cards — -20 °C to 70 °C
Humidity (non-condensing)
all cards — 40 °C to 80% RH
[1]
Not recommended for connection to ac line without external tran-
sient suppression. Up to 250 ch/s to internal memory. See scanning
rates for measurement condition and rate on each instrument.
[2]
Channel-to-channel or channel-to-earth
[3]
Errors included in DMM measurement accuracy specifi cations
[4]
50 Ω source, 50 Ω load
[5]
Bandwidth direct to card SMB connectors
[6]
Isolation within channel 1 to 20 or 21 to 40 banks is -40 dB
[7]
Applies to resistive loads only
[8]
Thermocouple measurements not recommended with 34908A
module due to common lo confi guration