beautypg.com

Xylem ProLab 2000 User Manual

Page 79

background image

ProLab 2000

Operation

79

ba75577e05

11/2013

5

Press

.
The measurement with stability control to determine the D.O.
concentration starts.

6

Wait for the end of the measurement with Stability control or
accept the measured value with

.
The determined D.O. content is displayed as a message (

i

).

7

Open the setting of the correction factor with

.

8

Using <

▲><▼>, set the correction factor to adjust the

displayed concentration value to the nominal value (value of
the comparison measurement). Subsequently, accept the
correction factor with

.

9

Select Accept with <

▲><▼> and press

.

The meter switches to the measured value display. The
[Factor] indicator appears in the measuring window.

[i] Concentration = 8.5 mg/l

[i] Temperature =24.8 °C

Terminate stability control

LFOx1400-ID

[i] Concentration = 8.49 mg/l

[i] Temperature =24.8 °C

Set factor:

1.000

Accept

LFOx1400-ID