What to do if...

103

ba75577e05

11/2013

Error message, OFL

Error message, Error

(invalid

calibration)

Error message, LEAK

Measured value flashes

for a long time

(No stable measured

value)

Measured value too low

Measured value too

high

Cause

Remedy

– Measuring range exceeded

– not possible

Cause

Remedy

D.O. sensor:

– Electrolyte solution depleted

– Regenerate probe

– Membrane contaminated

– Clean membrane

– Electrode system poisoned

– Regenerate probe

Cause

Remedy

D.O. sensor:

– Membrane damaged

– Exchange the membrane cap

– Membrane head screwed on

not tight enough

– Screw membrane head tight

Cause

Remedy

D.O. sensor:

– Membrane contaminated

– Clean membrane

Cause

Remedy

D.O. sensor:

– Insufficient flow

– Provide flow to the probe

Cause

Remedy

– High amount of dissolved

substances

– Correct solubility function

using the salinity equivalent

– Air bubbles bump on the

membrane with high velocity

– Avoid direct flow to the

membrane