Put the D.O. sensor into the air calibration vessel.

2

Connect the D.O. sensor to the meter.

3

In the measured value display, select the D.O. measuring
window with <

▲><▼> and .

4

Start the calibration with .
The calibration display for the relevant sensor type appears.

5

Put the D.O. sensor into the air calibration vessel.

6

Press

.
The measurement with stability control to determine the
relative slope starts.

7

Wait for the end of the measurement with Stability control or
accept the calibration value with

.
The determined relative slope is displayed as a message (

i

).

[i] Relative slope = 1.00

[i] Use air calibration vessel

Continue

LFOx1400-ID

[i] Relative slope = 1.00

[i] Temperature =24.8 °C

Terminate stability control

LFOx1400-ID