You can terminate prematurely the manual Stability control function
with manually at any time. If the manual Stability control
function is terminated prematurely, the current measurement data is
not downloaded to the interface.
Criteria
With identical measurement conditions, the following applies:
2
Freeze the measured value with
The display switches to the display of the selected channel as
necessary.
In the status line, [SC] is displayed.
3
With activate the Stability control function.
[SC] flashes while the stability control is active.
As soon as a stable measured value is recognized, the current
measurement data is downloaded to the interface. The
measured values of all channels meeting the criterion for
stability control are marked by SC.
4
Using
The [SC] status display disappears. The display switches to the
previous indication.
Measuring mode
Reproducibility
Response time
D.O. concentration
Better than 0.05 mg/l
> 10 seconds
D.O. saturation
Better than 0.6 %
> 10 seconds
D.O. partial
pressure
Better than 1.2 mbar
> 10 seconds