4 settings for conductivity measuring cells, Settings for conductivity measuring cells – Xylem ProLab 2000 User Manual
Page 66

Operation
ProLab 2000
66
ba75577e05
11/2013
4.7.4
Settings for conductivity measuring cells
Overview
The following settings are possible for conductivity measuring cells:
Measured parameter
Reference temperature
Temperature compensation
TDS factor
Calibration interval
Measuring cell / cell constant
Settings
The settings are made in the measuring menu of the conductivity
measurement. To open the settings, activate the relevant measuring
window in the measured value display and press the key.
After completing the settings, switch to the measured value display with
Menu item
Possible
setting
Description
Calibration /
Calibration interval
1 ... 999 d
Calibration interval for the
measuring cell (in days).
The meter reminds you to
calibrate regularly by the
flashing CalClock in the
measuring window.
Calibration /
Calibration record
-
Displays the calibration
record of the last calibration.
Calibration /
Calibration history
-
Displays the calibration
history of the last calibrations.
Reset
-
Resets all sensor settings to
the delivery condition (see
page 93).
Temp. comp. (TC) /
Reference temp.
20°C
25 °C
Reference temperature
This setting is only available
when the Conductivity or
Resistivity display is set.
Temp. comp. (TC) /
Compensation
nLF
lin
Off
Procedure for temperature
compensation (see page 65).
This setting is only available
when the Conductivity or
Resistivity display is set.