Xylem ProLab 2000 User Manual

Page 51

background image

ProLab 2000

Operation

51

ba75577e05

11/2013

Continuing with three-

point to five-point

calibration

(Buffer NIST/DIN)

12

Wait for the measurement with stability control to be completed
or Terminate stability control with

and accept the
calibration value.
The calibration display for the next buffer appears.

13

For two-point calibration, select Exit with 2 point calibration with
<

▲><▼> and confirm with

.

The calibration is completed as a two-point calibration.
The new calibration values are displayed as a message ([i]).

You have the following options:
 Accept the new calibration values with

.

Subsequently, the calibration record is displayed and
downloaded to the interface at the same time.

 To switch to the measured value display without accepting

the new calibration values, press or .

[i] Buffer value = 9.18

[i] U = -167.0 mV

[i] Temperature =

24.8 °C

Terminate stability control

pH

[i] Buffer series NIST/DIN

[i] Immerse sensor in buffer 3

Exit with 2 point calibration

Continue

pH

14

Thoroughly rinse the electrode with distilled water.

15

Immerse the electrode in buffer solution 3.

16

If necessary, measure the temperature of buffer 3 manually,
then enter and confirm it with <

▲><▼> and

in the

Set temperature setting.

17

Select Continue with <

▲><▼> and press

. The

buffer is measured.
The measured value is checked for stability (Stability control).