Xylem ProLab 2000 User Manual
Page 51

ProLab 2000
Operation
51
ba75577e05
11/2013
Continuing with three-
point to five-point
calibration
(Buffer NIST/DIN)
12
Wait for the measurement with stability control to be completed
or Terminate stability control with and accept the
calibration value.
The calibration display for the next buffer appears.
13
For two-point calibration, select Exit with 2 point calibration with
<
▲><▼> and confirm with .
The calibration is completed as a two-point calibration.
The new calibration values are displayed as a message ([i]).
You have the following options:
Accept the new calibration values with .
Subsequently, the calibration record is displayed and
downloaded to the interface at the same time.
To switch to the measured value display without accepting
the new calibration values, press
[i] Buffer value = 9.18
[i] U = -167.0 mV
[i] Temperature =
24.8 °C
Terminate stability control
pH
[i] Buffer series NIST/DIN
[i] Immerse sensor in buffer 3
Exit with 2 point calibration
Continue
pH
14
Thoroughly rinse the electrode with distilled water.
15
Immerse the electrode in buffer solution 3.
16
If necessary, measure the temperature of buffer 3 manually,
then enter and confirm it with <
▲><▼> and in the
Set temperature setting.
17
Select Continue with <
▲><▼> and press . The
buffer is measured.
The measured value is checked for stability (Stability control).