beautypg.com

3 accuracy (± 1 digit), 2 general data, Accuracy (± 1 digit) – YSI TruLab pH 1110 User Manual

Page 43

background image

TruLab pH 1110

Technical data

ba76155e01

03/2014

43

11.1.3

Accuracy (± 1 digit)

11.2

General data

Parameter

Accuracy

Temperature of the

test sample

pH / range

*

- 2.0 ... + 20.0
- 2.00 ... + 20.00
- 2.000 ... + 19.999

* when measuring in a range of ± 2 pH around a calibration point

± 0.1
± 0.01
± 0.005

+ 15 °C ... + 35 °C
+ 15 °C ... + 35 °C
+ 15 °C ... + 35 °C

U [mV] / range
- 2000 ... + 2000
-1200.0 ... +1200.0

± 1
± 0.3

+ 15 °C ... + 35 °C
+ 15 °C ... + 35 °C

T [°C] / temperature sensor
• NTC 30
• PT 1000

± 0.1
± 0.1

Dimensions

Approx. 230 x 190 x 80 mm

Weight

Approx. 1.0 kg

Mechanical
structure

Type of protection IP 43

Electrical
safety

Protective class III

Test certifi-
cates

CE, cETLus

Ambient
conditions

Storing: - 25 °C ... + 65 °C

Operation: +5 °C ... + 55 °C
With the power pack connected:
+5 °C ... + 40 °C

Allowable relative humidity
Yearly mean: < 75 %
30 days/year: 95 %
Other days: 85 %