03/2014

6.1.2

Measuring the temperature

The temperature measurement is absolutely essential for a reproducible ORP
measurement.

You have the following options to measure the temperature:

• Measurement with an external temperature sensor.
• Manual determination and input of the temperature.

The measuring instrument recognizes whether a suitable sensor is connected
and automatically switches on the temperature measurement.

Which type of temperature measurement is active is indicated by the tempera-
ture display and the [TP] status indicator:

If the measurement is made without a temperature sensor, proceed as follows:

1.

Measure the current temperature of the test sample.

2.

Set the temperature value with <><>.

6.2

Calibration

ORP electrodes are not calibrated. You can, however, check ORP electrodes
using a test solution.

Temperature

sensor

Resolution of the

temp. display

Temp. measurement

yes

0.1

Automatic with temperature sensor

-

1 °C

Manual