8 reset, 1 resetting the calibration values, 2 resetting the measurement and system settings – YSI TruLab pH 1110 User Manual
Page 34: Reset, Resetting the calibration values, Resetting the measurement and system settings, 8reset
Reset
TruLab pH 1110
34
ba76155e01
03/2014
8
Reset
You can erase the calibration values and reset (initialize) the measurement and
system settings.
8.1
Resetting the calibration values
1.
Press
InIt.C is displayed.
2.
Use <><> to display no or YES.
•
YES: Reset the calibration values
•
no: Retain the calibration values.
3.
Confirm with
Calibration values to be reset
8.2
Resetting the measurement and system settings
1.
Switch on the meter with
The display test appears briefly on the display.
2.
During the display test, press
meter settings.
Init is displayed.
3.
Use <><> to display no or YES.
•
YES: Reset the meter settings.
•
no: Retain the meter settings.
4.
Confirm with
The meter switches to the measuring mode.
The calibration values are reset to default. All other meter
settings are retained.
Recalibrate after performing a reset.
Calibration value
Default settings
Zero point (ASY )
0 mV (pH 7.000)
Slope (SLO )
-59.16 mV/pH (100 %)