TruLab pH 1110
pH value
ba76155e01
03/2014
23
3.
If necessary, press
ASY is displayed.
4.
Thoroughly rinse the pH electrode with deionized water.
5.
Immerse the pH electrode in the first buffer solution (pH 7.0 ± 0.5 for
two-point calibration).
6.
When measuring without temperature sensor: Enter the temperature of
the buffer with <><>.
7.
Start the measurement with
The pH value of the buffer solution is displayed.
The measured value is checked for stability (stability control).
The [AR] status display flashes.
8.
Wait for the measurement with stability control to be completed.
9.
Set the nominal pH value of the buffer solution with <><>.
10. Accept the calibration value with
SLO is displayed.
11. If necessary, finish the calibration procedure as a single-point calibration
with
The calibration record is displayed.
or
Continue calibration using the next buffer with
Continuing with two-point calibration
12. Thoroughly rinse the pH electrode with deionized water.
13. Immerse the pH electrode in the second buffer solution.