Calibration determines and stores the current values of the zero point and
slope. Thus, you should calibrate at regular intervals.
5.2.2
When to calibrate?
• When the calibration interval has expired
• Routinely within the framework of the company quality assurance
• After connecting another combination electrode
5.2.3
Automatic calibration (AutoCal)
Use any one to three buffer solutions of the selected buffer set in ascending or
descending order.
Below, calibration with YSI buffers (TEC) is described. When other buffer sets
are used, other nominal buffer values are displayed. Apart from that, the proce-
dure is identical.
1.
Start the calibration with
The calibration display for the first buffer appears.
2.
When measuring without temperature sensor: Temper the buffer or
measure the current temperature.
3.
If necessary, press
[AutoCal DIN]).
Ct1 or Cd1 is displayed.
4.
Thoroughly rinse the pH electrode with deionized water.
5.
Immerse the pH electrode in the first buffer solution.
6.
When measuring without temperature sensor: Enter the temperature of
the buffer with <><>.
7.
Start the measurement with
The measured value is checked for stability (stability control).
The [AR] status display flashes.
The electrode voltage (mV) or the nominal value of the buffer is displayed
(setting: see section 7.1.1 C