4 manual calibration (concal), Manual calibration (concal), 3t c – YSI TruLab pH 1110 User Manual

Page 22

background image

pH value

TruLab pH 1110

22

ba76155e01

03/2014

15. If necessary, finish the calibration procedure as a two-point calibration

with .
The calibration record is displayed.

or
Continue calibration using the next buffer with .

Continuing with three-point calibration

16. Thoroughly rinse the pH electrode with deionized water.
17. Immerse the pH electrode in the third buffer solution.
18. When measuring without temperature sensor: Enter the temperature of

the buffer with <><>.

19. Start the measurement with .

The measured value is checked for stability (stability control).
The [AR] status display flashes.
The electrode voltage (mV) or the nominal value of the buffer (setting: see
section 7.1.1 C

HANGING

THE

SETTINGS

FOR

P

H

MEASUREMENTS

, page 29).

20. Wait for the measurement with stability control to be completed or termi-

nate the stability control and take over the calibration value with
.
The calibration record is displayed.

5.2.4

Manual calibration (ConCal)

Use any buffer solution for the single-point calibration. The calibration will be
the more accurate the nearer the pH value of the buffer solution is to that of the
test sample.

Use the following buffer solutions for two-point calibration:

• one buffer solution with pH 7.0 ± 0.5
• any other buffer solution

1.

Start the calibration with .
The calibration display for the first buffer appears.

2.

When measuring without temperature sensor: Temper the buffer or
measure the current temperature.

8

4

2

3

T

C

°C

pH

AutoCal TEC

TP