Page 24: Calibration data

background image

pH value

TruLab pH 1110

24

ba76155e01

03/2014

14. When measuring without temperature sensor: Enter the temperature of

the buffer with <><>.

15. Start the measurement with .

The pH value of the buffer solution is displayed.
The measured value is checked for stability (stability control).
The [AR] status display flashes.

16. Wait for the measurement with stability control to be completed.
17. Set the nominal pH value of the buffer solution with <><>.
18. Accept the calibration value with .

The calibration record is displayed.

5.2.5

Calibration points

Depending on the number of buffer solutions used, the meter determines the
following values and calculates the calibration line:

5.2.6

Calibration data

Displaying the calibration data

1.

Display the calibration data in the measured value display with .
The value for the asymmetry (ASY) is displayed.

Calibration

Values deter-

mined

Displayed calibration data

1-point

Asy

• Zero point = Asy
• Slope = Nernst slope

(-59.2 mV/pH at 25 °C)

2-point

Asy
Slo

• Zero point = Asy
• Slope = Slo
The calibration line goes through both cal-
ibration points.

3-point

Asy
Slo

• Zero point = Asy
• Slope = Slo
The calibration line is calculated by linear
regression.

You can display the slope in the units, mV/pH or % (see section
5.2.6 C

ALIBRATION

DATA

, page 24).