10 what to do if, 1 ph, 1 no stable measured value – YSI TruLab pH 1110 User Manual
Page 38: 2 error message calerror, What to do if, No stable measured value, Error message calerror, See section 10 w
What to do if...
TruLab pH 1110
38
ba76155e01
03/2014
10
What to do if...
10.1
pH
10.1.1
No stable measured value
10.1.2
Error message CalError
More information and instructions on cleaning and exchange of
sensors are given in the documentation of your sensor.
Cause
Remedy
• Junction of the electrode contami-
nated
• Clean the junction
• Glass membrane of the electrode
contaminated
• Clean the glass membrane
Cause
Remedy
• pH value of the test sample not
stable
• Measure with air excluded if
necessary
• Temperature of the test sample
not stable
• Adjust temperature if necessary
Cause
Remedy
• The values determined for zero
point and slope of the electrode
are outside the allowed limits.
• Recalibrate
• Junction contaminated
• Clean the junction
• pH electrode broken
• Replace the pH electrode
- 3100 (50 pages)
- 3200 (86 pages)
- 5000 (70 pages)
- 5200 (148 pages)
- 5200A (128 pages)
- 5200A (121 pages)
- 5400 (127 pages)
- 5239 Dissolved Oxygen Probe (23 pages)
- 55 (26 pages)
- 550A (18 pages)
- 556 MPS - Quick Start Guide (4 pages)
- 58 (40 pages)
- 5905 (22 pages)
- 6600V2 - User Manual (379 pages)
- 60 (33 pages)
- 63 (46 pages)
- 650 MDS (72 pages)
- 9500 Photometer (143 pages)
- 9300 Photometer (11 pages)
- 9500 Photometer (18 pages)
- CR 3200 Thermoreactor (42 pages)
- CR 4200 Thermoreactor (42 pages)
- CR2200 Thermoreactor (36 pages)
- DO200 (40 pages)
- EC300 (40 pages)
- EC300A (40 pages)
- EC30A (12 pages)
- IFL 700 IQ (46 pages)
- IQ SensorNet 2020 XT Terminal (196 pages)
- IQ SensorNet FDO Sensor (48 pages)
- IQ SensorNet MIQ/A-MOD Module (42 pages)
- IQ SensorNet MIQ/A-PR Module (36 pages)
- IQ SensorNet MIQ/IC2 Module (30 pages)
- IQ SensorNet MIQ/IF232 Module (18 pages)
- IQ SensorNet MIQ/JB Module (18 pages)
- IQ SensorNet MIQ/JBR Module (20 pages)
- IQ SensorNet MIQ/PS Module (22 pages)
- IQ SensorNet MIQ/R6 Module (56 pages)
- IQ SensorNet NitraLyt Sensor (64 pages)
- IQ SensorNet SensoLyt Sensor 700 IQ (SW) (46 pages)
- IQ SensorNet Sensors Encoded Fieldbus Data (12 pages)
- IQ SensorNet TetraCon Sensor 700 IQ (SW) (32 pages)
- IQ SensorNet ViSolid Sensor 700 IQ (60 pages)
- MultiLab 4010-1 (108 pages)
- MultiLab 4010-2 (138 pages)