What to do if...

TruLab pH 1110

38

ba76155e01

03/2014

10

What to do if...

10.1

pH

10.1.1

No stable measured value

10.1.2

Error message CalError

More information and instructions on cleaning and exchange of
sensors are given in the documentation of your sensor.

Cause

Remedy

• Junction of the electrode contami-

nated

• Clean the junction

• Glass membrane of the electrode

contaminated

• Clean the glass membrane

Cause

Remedy

• pH value of the test sample not

stable

• Measure with air excluded if

necessary

• Temperature of the test sample

not stable

• Adjust temperature if necessary

Cause

Remedy

• The values determined for zero

point and slope of the electrode
are outside the allowed limits.

• Recalibrate

• Junction contaminated

• Clean the junction

• pH electrode broken

• Replace the pH electrode