1 recorder – YSI IQ SensorNet DIQ/CR3 User Manual
Page 44
Settings
MIQ/CR3; DIQ/CR3
4 - 30
ba76032e01
01/2012
4.6.1
Recorder
Function
The measured values of the linked sensor at the current output are set
up as current intensity in the Recorder application. The output of
measured values is laid down in the Recorder type, Start value and End
value settings.
Settings
Function
Settings
No function
The current output is not used.
Recorder
PID controller
Fixed current value
8
Carry out the settings for the current output with
d
and
g
.
9
Highlight and confirm Save and quit with
d
and
g
.
The new settings are stored.
Setting
Selection/Values
Explanation
Recorder type
0 to 20 mA or
4 to 20 mA
Start value
within the
measuring range
(sensor-
dependent)
Minimum spacing:
5 % of the measuring
range
End value
Measured variable
Main variable
Adjoining variable
Main variable designates
the actual measured
parameter of the sensor
(e.g. pH, oxygen, etc.).
Adjoining variable
designates an additional
measured parameter
(e.g. temperature).
Attenuation
0 ... 40 mA/s
Speed of change of the
output current (mA/s) in
the case of an erratic
change of the input
signal.