Warning, Connecting lines to the terminal strip, Miq/cr3; diq/cr3 installation 3 - 5 – YSI IQ SensorNet DIQ/CR3 User Manual

Page 13

background image

MIQ/CR3; DIQ/CR3

Installation

3 - 5

ba76032e01

01/2012

Connecting lines to the

terminal strip

Fig. 3-1

Terminal strip with the relay and current connections

Warning

No free wires must be allowed to project into the enclosure.
Otherwise, there is a danger that safe areas could come into
contact with dangerous voltages. This could result in life
threatening electric shock when working with the IQ S

ENSOR

N

ET

.

Always cut off any wires that are not in use as closely as possible
to the cable gland.

1

Open the module.

2

Screw the cable gland (pos. 1 in Fig. 3-1) with the sealing ring
(pos. 2) into the module housing.

3

Loosen the coupling ring (pos. 3 in Fig. 3-1).

4

Feed the line through the cable gland in the module housing.

5

Connect the wires to the terminal strip. While doing so, pay
attention to the specifications on the label located under the
terminal strip.

6

Tighten the coupling ring (pos. 3 in Fig. 3-1).

1
3

2

Relay

Current output

X18

X17

250 VAC

5 A AC

R1

X16

X15

250 VAC

5 A AC

R2

X14

X13

250 VAC

5 A AC

R3

X12

X10

X8

X11

X9

X7

0/4...20mA 0/4...20mA 0/4...20mA

+ REC -

C1

+ REC -

C2

+ REC -

C3

X6

X5

X4

SENSORNET 2

RED

SHIELD

GREEN

X3

X2

X1

SENSORNET 1

RED

SHIELD

GREEN

ON

OFF

SN

TERMINA

TO
R