Warning, Connecting lines to the terminal strip, Miq/cr3; diq/cr3 installation 3 - 5 – YSI IQ SensorNet DIQ/CR3 User Manual
Page 13
MIQ/CR3; DIQ/CR3
Installation
3 - 5
ba76032e01
01/2012
Connecting lines to the
terminal strip
Fig. 3-1
Terminal strip with the relay and current connections
Warning
No free wires must be allowed to project into the enclosure.
Otherwise, there is a danger that safe areas could come into
contact with dangerous voltages. This could result in life
threatening electric shock when working with the IQ S
ENSOR
N
ET
.
Always cut off any wires that are not in use as closely as possible
to the cable gland.
1
Open the module.
2
Screw the cable gland (pos. 1 in Fig. 3-1) with the sealing ring
(pos. 2) into the module housing.
3
Loosen the coupling ring (pos. 3 in Fig. 3-1).
4
Feed the line through the cable gland in the module housing.
5
Connect the wires to the terminal strip. While doing so, pay
attention to the specifications on the label located under the
terminal strip.
6
Tighten the coupling ring (pos. 3 in Fig. 3-1).
1
3
2
Relay
Current output
X18
X17
250 VAC
5 A AC
R1
X16
X15
250 VAC
5 A AC
R2
X14
X13
250 VAC
5 A AC
R3
X12
X10
X8
X11
X9
X7
0/4...20mA 0/4...20mA 0/4...20mA
+ REC -
C1
+ REC -
C2
+ REC -
C3
X6
X5
X4
SENSORNET 2
RED
SHIELD
GREEN
X3
X2
X1
SENSORNET 1
RED
SHIELD
GREEN
ON
OFF
SN
TERMINA
TO
R