YSI IQ SensorNet DIQ/CR3 User Manual
Page 39

MIQ/CR3; DIQ/CR3
Settings
4 - 25
ba76032e01
01/2012
Example
Fig. 4-16 Cleaning cycle
Setting
Result
Reference time (h):
Reference time
(min):
Interval unit:
Cleaning interval:
12
0
Hours (h)
8 h
Reference time: 12:00 hours
This specifies the following
start times:
04:00, 12:00 and 20:00 hours
1
Reference time
Start of a cleaning cycle (t1)
Start of the specified Cleaning duration (t1a)
2
End of the specified Cleaning duration (t1a)
Start of the specified Adjustment time (t1b)
3
End of the specified Adjustment time (t1b)
End of the cleaning cycle (t1)
4
Reference time ± Cleaning interval (t2)
Start of a cleaning cycle
t1
Cleaning cycle =
Cleaning duration (t1a) Adjustment time (t1b)
Linked outputs are frozen.
t2
Cleaning interval
relay
condition
open
closed
00:00
Time
3
1
2
t1
t1
t1
t1a t1b
4
4
t2
t2
t2
t2
1
04:00
12:00
24:00
20:00