3 proportional output, Proportional output -5, See section 4.1.3 – YSI IQ SensorNet DIQ/CR3 User Manual
Page 19
MIQ/CR3; DIQ/CR3
Settings
4 - 5
ba76032e01
01/2012
4.1.3
Proportional output
In the case of proportional output, a relay switches cyclically on and off
in a defined measured value range (proportional range). At the same
time, the relay switches with a:
duration of operation that corresponds to the measured value
(pulse-width output, see page 4-7) or
switching frequency (frequency output, see page 4-8).
Proportional outputs can be used in the following way:
Output with one relay:
An output range is defined with a Start value and an End value. No
output takes place above and below the output range
(see page 4-6).
Output with two relays:
An output range is defined for each relay with a Start value and an
End value. One relay outputs in the upper output range and a further
relay in the lower output range (see page 4-6).