6504, front cover removed, 2) black common (3) purple sdi-12 – YSI 6500 Environmental Process Monitor User Manual
Page 87

Accessories Appendix
D
Enter
Esc
Cal
6500
6500
EN VIRON MEN TAL
EN VIRON MEN TAL
MON ITORIN G
MON ITORIN G
SYSTEM
SYSTEM
21.2 Temp
7.35 DO
6.53 pH
6507
6’ Patch cable
w/ MS-8
Sonde cables
w/ MS-8
Stream
Sondes
6504
Breakout Box
6507
Gland Fitting
Enter
Esc
Cal
6500
6500
EN VIRO N MEN TAL
EN VIRO N MEN TAL
MO N ITO RIN G
MO N ITO RIN G
SYSTE M
SYSTE M
21.2 Temp
7.35 DO
6.53 pH
Junction Box
6508
Sonde cables
w/ MS-8
Stream
Sondes
Breakout Box
Conduit &
6507
6’ Patch cable
w/ MS-8
customer-supplied
3-conductor cable
In remote Breakout Box installations, Figure 4, the 6500 Monitor uses a standard MS-8
connection to interface with the 6 foot (1.8 m) 6507 Patch Cable that is connected in the 6508
Junction Box. The Junction Box is then connected by conduit and three- conductor cable to the
Breakout Box(s). The customer is advised to supply rigid conduit and 18 AWG or heavier
shielded multi-conductor cable to connect between the Junction Box and the Breakout Box at the
Sonde installation site. This configuration allows a maximum cable length of 250’ between the
6500 Monitor and the Sonde(s). The cable from the junction box should be landed to TB-1 in the
breakout box. The MS-8 connectors for Sonde hookups are pre-wired to TB-2 and TB-3. See
Figure 12 for further information.
6504, Front Cover Removed
(2) MS-8 to Sonde
3/4” conduit to Jct. Box
The TB-1, Figure 12, conductor color
code is:
(1) Red
+12 VDC
(2) Black
Common
(3) Blue
SDI-12
IMPORTANT!
GND on TB-1 is for signal common only.
Do not connect to earth ground.
1
3
2
TB-1
Figure
11
Figure 10
The TB-2 & TB-3, Figure 12, conductor color code is:
(1) Red
+12 VDC
Figure12
(2) Black
Common
(3) Purple
SDI-12
YSI Incorporated
6500 Environmental Process Monitor
84