3.3.1 CALIBRATION SETUP
Highlight the Calibration setup selection and press Enter. The following screen will be displayed.
The Calibration Setup submenu presents two items:
(1) Calibrations enabled is a list of all possible sensor calibrations that can either be enabled or
disabled with regard to calibration from the Run mode.
(2) Advanced setup is a selection that allows the user to choose whether calibration errors can be
overridden and defines the default calibration value that will appear on the display during the
calibration protocol.
Highlight the Calibration enabled selection, press Enter, and the following screen will be
displayed.
Once a calibration is selected as indicated by the check mark, that sensor may be calibrated using
the Cal key during the Run mode. The user can select and deselect items using the
↑↓ arrows to
highlight and press Enter to toggle selections on and off.
In the example shown above, the parameters specific conductance, DO % air saturation, DO mg/L
(dissolved oxygen in mg/L concentration units) can be calibrated during the Run mode of the
6500. This selection will NOT allow calibration of conductivity or salinity even if these readings
appear on the display.
Press Esc to return to the Calibrate setup menu.