Installation Section

2

The 6507 Patch Cable is pre-wired. The conductor color code is: Red

+12 VDC

Black Common

Purple SDI-12


6502, Front Cover Removed

MS-8 to sonde

3/4” conduit to jct box


1

3

2

+1

2

VD

C

GN

D

SDI

-12

TB-1

Figure 8
Connector TB-1
detail




6508

Enter

Esc

Cal

6500

6500

EN VIRO N MEN TAL

EN VIRO N MEN TAL

MO N ITO RIN G

MO N ITO RIN G

SYSTEM

SYSTEM

21.2 Temp
7.35 DO
6.53 pH

Junction

Box

sonde cable

w/ MS-8

Stream

Sonde

Breakout

up to 250 ‘

conduit

6507

6’ Patch cable

w/ MS-8

customer-supplied

3-conductor cable

1

3

2

+12
VDC

GND

SDI-
12

TB-1




IMPORTANT!

GND on TB-1 is for signal common only. Do not connect to earth ground.

YSI Incorporated

6500 Environmental Process Monitor

12