beautypg.com

6 q-factor calculation, 7 load supervision applications – Crown Audio IQ-PIP-USP2 User Manual

Page 18

background image

IQ-PIP-USP2

Page 35

IQ-PIP-USP2 Reference Manual

IQ-PIP-USP2

Page 34

IQ-PIP-USP2 Reference Manual

5.6 Q-Factor Calculation

Many of the DSP filters on the IQ-
PIP-USP2 feature adjustable Q-fac-
tor, which is a measure of the filter’s
selectivity or sharpness. Q-factor is
adjustable in fractions of an octave
on the IQ-PIP-USP2.

Use the table in Figure 5.9 to deter-
mine Q-factor for a given bandwidth

in octaves. The table relates the Q-
factor to bandwidth in octaves ac-
cording to the equation:

where K is the number of equal divi-
sions per octave.

Figure 5.9 Q-factor vs.Bandwidth

2

(1/2K)

1

(1/K)

–1

Q=

Q-factor

0.200

0.266

0.667

1.414

2.415

2.871

4.318

8.651

14.42

28.85

43.28

50.00

100.0

K

0.21

0.25

0.5

1

1.5

2

3

6

10

20

30

34.7

69.3

BW in Octaves

4.8

4

2

1

2/3

1/2

1/3

1/6

1/10

1/20

1/30

1/35

1/70

5.7 Load Supervision
Applications

The IQ-PIP-USP2 Load Supervision
feature can be used to monitor the
amplifier load in real time with al-
most any program material. Aver-
age load impedance is calculated
as a function of amplifier output volt-
age and current. The system re-
quires approximately 20-30 mA of
average amplifier output current for
adequate supervision. This allows
typically low average output power
levels of less than ½ watt with most
loads. The maximum load imped-
ance for reliable system perfor-
mance is limited to about 50 ohms.
Higher impedances can be mea-
sured but may require higher ampli-
fier output levels for reliable opera-
tion.

Most amplifier/load systems can be
configured and supervised by fol-
lowing these steps:

1 Configure your audio system

using a known “good” load, then
enable the Load Supervision
feature.

2 Provide typical program material

at a level high enough to light the
“test” indicator.

3 Run the system at this level until the

average impedance stabilizes.
This may take seconds to minutes
depending on level, duty-cycle,
etc.

4 Set the nominal impedance at the

measured value average. This
optimizes the supervision
algorithm for voltage and current
levels versus the actual load Note:
a higher nominal setting will
require higher output levels.

5 Set the high limit at twice average

and the low limit at one-fourth
nominal.*

6 Let the system run for extended

periods using any and all typical
program material.

7 Adjust the high/low limits, if

necessary, to account for any
variance in average measured
impedance.

8 Enable error reporting, if desired.

This procedure should work well for
most applications. However, some
applications can be a little more
difficult. Some very low-level and/or
low duty-cycle signals may not ad-
equately “test” the load. Lab and
situation testing have shown output
levels as small 40 dB below rated
amplifier output to be enough for
most low-impedance loads. Higher
impedance loads such as those used
in “lightly-loaded” 70V distribution
lines may require signal level near
20 dB below rated output.

The “Nominal Load Impedance” con-
trol is used to optimize the system
for the most accurate calculation of
load impedance. It should be set to
the expected nominal (or rated) im-
pedance of the “normal” load. The
high limit should be set for at least 2
times the expected nominal or ac-
tual measured load, while the low
limit should be set to ½ the expected
nominal or actual measured load.

* These limits are somewhat arbitrary but

should be a good starting point.